Recommended separation distances between portable and mobile RF communications equipment and the
HFA II
-i
The HFA II
-i
is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the HFA II
-i
can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the HFA II
-i
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d = 1.17
P
d = 1.17
P
d = 2.33
P
0,01
0.117
0.117
0.233
0,1
0.370
0.370
0.737
1
1.170
1.170
2.330
10
3.700
3.700
7.368
100
11.700
11.700
23.300
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Introduction & Instrument Setup
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
1-16
Accessory Equipment
WARNING: Accessory equipment connected to the analog and digital interfaces must
be certified according to the respective IEC standards (e.g., IEC 60950 for data
processing equipment and IEC 60601-1 for medical equipment). Furthermore, all
configurations shall continue to comply with the applicable system requirements of
IEC 60601-1 standard
.
Any person who connects additional equipment to the signal
input part or signal output part configures a medical system, and is therefore
responsible for ensuring that the complete system complies with the applicable
system requirement of IEC 60601-1 standard. If in doubt, consult the technical service
department or your local representative.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......