5 First Operating Steps | 5.7 Working with Specific Specimen Types
ZEISS
15. Adjust the working parameters, refer to
.
16. To return to HV mode, repeat steps 1 to 5 and step 7 to dismount the aperture, deselect
the aperture in the
Apertures
tab and click
Go To HV
in the
Vacuum
tab.
5.7.2.2 Aligning the Gun and Mid-column Aperture for EasyVP Mode
To achieve an optimal alignment of the beam through the 20 μm mid-column aperture, you need
to perform an alignment procedure.
Info
You can refer to
Using the Automatic Gun Alignment Functions [
Info
Once the alignment is optimized, do not change the position of the X and Y micrometer
gauges on the mid-column aperture changer.
Prerequisite
ü
The EasyVP aperture is fitted.
ü
The 20 µm aperture is selected at the mid-column aperture changer (position 2).
Procedure
1. From the
Menu Bar
, select
Tools > User Preferences
.
à
The
User Preferences
dialog is displayed.
2. Check that
Auto Calibration
and
User Align
are set to
Yes
.
3. Adjust the following settings in the
SEM Controls
panel:
Select the EasyVP Aperture
Optibeam = Resolution
Aperture Size = 20.00 µm
EHT Target = 20.00 kV
I Probe = 100 pA
WD = 8.5 mm
Signal A = SE1
Operating Mode = Normal
INFO:
For details on where to adjust these settings, refer to
.
4. To carry out a C1/C2/C3 lens hysteresis removal, press
<Shift + F2>
.
5. To create an emission image, select
Beam > Gun Align
from the
Menu Bar
.
à
The
Gun Alignment
window is displayed.
96
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006