9 Technical Data and Conformity | 9.3 Performance Data and Specifications | EVO 10
Parameter
Value
8 mm) which is directly connected to the PE
of the FEMALE RECEPTACLE as short as possi-
ble.
This grounding connection must not be com-
mon to other electrical equipment.
A grounding wire AWG10 (≥ 5 m) is delivered
with the microscope. It serves as connection
between the grounding screw terminal (Ø
8 mm) and the microscope (Ø 6 mm).
Cross section
: Min. AWG10 (between
grounding screw terminal and PE of the CEE
FEMALE RECEPTACLE)
Circuit breaker type (at house installation)
16 A
Type C
Protection class
Class I
Momentary interruption
Less than a half cycle
Ampere interrupting capacity (AIC)
Min. 10,000 A
rms
Location
Requirements
Parameter
Requirement
Installation site
Exclusively inside buildings
Recommended
room size
Min. 3.6 m × 4.0 m × 2.3 m
Service area
Min. 0.8 m at each side
Entrance
Min. 0.8 m wide
Hallways
Min. 1.0 m wide
Corners
Min. 1.2 m wide
Transport ways
Free of staircases
Installation cate-
gory
II
Floor stability
> 1000 kg/m²
Exhaust Line
If toxic chemicals or biological specimens are used an exhaust line is recommended to remove the
waste gas of the pre-vacuum pump and to transmit it to the outside.
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006
155