Tick the checkbox to activate online stitching. After
ticking the checkbox the
Threshold
input box
becomes available. Enter the desired threshold
in
%
.
By pressing
Add
you can mark stage positions,
which will be listed in the
Marked position
field
in terms of Number as well as
x [
µ
m]
and
y [
µ
m]
coordinates (Fig. 202). By Pressing
Remove
the
highlighted position will be removed from the list.
Pressing
Remove all
will delete all positions. You
can save a position list by Pressing the
Save…
button in the
Mark position
field via the
Windows Explorer. You can load saved position list
by pressing the
Load…
button via selection in
Windows Explorer.
If
Centered grid
is selected, the current position
will be the center of the Tile Scan.
If
Bounding grid
is selected positions that should
be part of the tile scan have to be marked using
Add
. To this end click on the edge of the square to
select the position, which will then be displayed
with a thicker line width, and then press
Add
.
Using these positions a bounding grid is created,
which finally defines the dimensions of the tile
scan (Fig. 203).
If
Convex hull
is selected positions that should be
part of the tile scan have to be marked using
Add
.
To this end click on the edge of the square to
select the position, which will then be displayed
with a thicker line width, and then press
Add
.
Using these positions a convex hull is created,
which finally defines the dimensions of the tile
scan (Fig. 204).
Marked positions
are listed up (Fig. 205). This list
can be saved and loaded.
Fig. 203
Tile Scan tool – Bounding grid
selected; two positions defined
Fig. 204
Tile Scan tool – Convex hull selected
Fig. 205
Tile Scan: Marked positions