START-UP
DirectFRAP
Connecting the Fiber Oscillator
Carl Zeiss
05/2014
423635-9001-701
41
3.4
Connecting the Fiber Oscillator
The fiber oscillator makes the optical fiber oscillate, providing homogeneous bleaching or
photomanipulating of the ROI.
•
Place the fiber oscillator separately from the imaging system for laser manipulation, e.g. on the floor; it
must not stand on the vibration-damped table.
•
Clamp the fiber into the holding clip (Fig. 3-3/
1
).
•
Connect the CAN connector (Fig. 3-3/
2
) of the fiber oscillator with a free CAN port on the microscope.
•
Connect the trigger connector (Fig. 3-3/
3
) with the signal distribution box SVB (generic 1;
"digital out 7" or "digital out 8").
•
The fiber oscillator is switched on (5 V) or off (0 V) only via the trigger connector.
1
Holding clip
3
Trigger connector
2
CAN connector
Fig. 3-3
Connecting the fiber oscillator
Summary of Contents for DirectFRAP
Page 1: ...DirectFRAP Operating Manual...