3.5. Technical data
3.5.1. Specifications
Performance
SEM Resolution
High resolution configuration
(max.probe current 20 nA)
Optimum Working distance:
0.9 nm at 30 kV (STEM mode)
1.0 nm at 15 kV
1.9 nm at 1 kV
Acceleration voltage
Range: 0.02 kV - 30 kV (SEM)
Adjustment:
Continuously variable in 10 Volt steps.
Probe Current
SEM
5 pA to 20 nA High resolution option
12 pA to 100 nA High Current option
Magnification
Range: 12x – 2.000.000x (SEM)
Adjustment
: Continuously variable in either coarse or fine modes.
Auto-Compensation
: Magnification is precisely corrected
automatically for changes in working distance or acceleration
voltage.
Electron optics
Electron source
Filament
: Schottky field emitter.
Alignment
: Two stage electromagnetic emitter alignment utilizing emission
imaging mode
Automatic Emitter Run-Up
: Safe controlled run-up of the target emitter
conditions
Indicators
: Status information in Software to indicate all emitter parameters
Lens control
Patented GEMINI I VP electromagnetic/electrostatic objective lens system
(68º conical final-lens) with water cooling for best thermal stability and
reproducibility
Hysteresis Correction
: Built-in, for the removal of lens hysteresis
Summary of Contents for Crossbeam 340
Page 1: ...Crossbeam 340 Crossbeam workstation Instruction Manual ...
Page 24: ...OQ çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF OK p ÑÉíó p ÑÉíó ÉèìáéãÉåí ...
Page 60: ...SM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF RK fåëí ää íáçå ...
Page 120: ...NOM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Page 126: ...NOS çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF NNK ÄÄêÉîá íáçåë ...
Page 130: ...NPM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...
Page 133: ......