45
Corona process | User Manual
Maintenance
Corona process
7.3 Lamp calibration
•
InProcess software V 2.7.0
is installed on your PC.
•
You have replaced both halogen lamps and the measuring system has been
properly closed again.
•
Have the following tools and accessories ready
– Corona process calibration bench, order no. 000000-2211-998
Scope of supply: � Calibration device
� Certified white standard with appropriate certificate
(attached CD)
� Black standard
� Tools for changing lamps
� Corresponding instructions
� Storage box
– Ethernet cable, order no. 000000-2036-573
– Lab power supply, order no. 000000-2045-548
1
Calibration device setup
Fold open the calibration device.
Have the reference standard ready.
2
Preparing the measuring system for calibration
Mount the
Corona process
on the calibration device.
Fig. 26 Mounting the Corona process on the calibration device
Connect the Ethernet cable and the power supply cable to the measuring
system.
Requirements
Procedure