Installing the measuring system
Corona process
4 Installing the measuring system
4.1 Preparing the installation
1
Compare the delivery documents with the system components actually
supplied to you.
2
Check all system components for external intactness.
3
Check the existing environmental and connection conditions. The following
values must be observed:
Supply voltage
12 – 24 V
SELV
(safety extra low voltage)
Power consumption
40 W
Inrush current
< 4 A
Operating temperature
–10 to + 50 °C
Storage/transport tempera-
ture
– 40 to + 70 °C
Max. humidity
95 % non-condensing
Altitude of operation
Up to 2000 m
Tab. 2 Connection and environmental conditions
4
If you do not use the original mounting device from Carl Zeiss Spectroscopy
GmbH (
000000-2107-066
), make sure that the customer-installed mechanical
interface has the dimensions required to mount the
Corona process.
The device must be stable enough to support the weight of the
Corona process
(15 kg) vibration-free.
Note
The exact connection dimensions are specified in Section 3.2.4
“Mechanical interface” on page 17.
5
Make sure that all cables can be laid so that they are protected and free of
mechanical and thermal loads.
Procedure