Service manual | Lamp replacement and calibration - Corona process
Replacing halogen lamps
Lamp replacement and calibration Corona process
3
Reinstall the device base to the cover
(see Fig. 15 – reversed)
Caution
RISK OF COMPONENT DAMAGE
Risk of damaging the FBG filter board whilst placing the base into the
cover.
Carefully place the base and watch out for stuck or jammed parts and
cables.
Carefully place the base of the device
2
into the device cover
4
by
holding it on the handle
3
.
Attach the device base to the device cover by using the four hexagon bolts
1
.
4
Calibrating the compensation data
After each lamp replacement, the
Corona Process
measuring system must be
calibrated (see Chapter „5 Sensor calibration“).
Concluding measures