Checking the safety devices
61211-1310402
CONTURA
2-17
v
x
F
v
X
:
Ram speed in the X axis
F:
Force which must be exerted manually against the housing cover of the
ram.
Under normal circumstances, the force of your hand should suffice to
stop movement. If the movement cannot be stopped in this way, the
protective circuit must be checked immediately by a ZEISS service engi-
neer.
NOTE
If you can stop the travel movements, you will need to reactivate the
drives. See control elements.
Y-Axis
Components may be damaged by improper handling.
The probe, ram, and left column could be damaged.
ü
You may only push against the right column of the bridge.
Do
not
push against the probe and do
not
push against the left col-
umn of the bridge.
1
Move the control console to the
drive side
for Y axis travel.
The drive is located on the
right
side of the CMM.
Summary of Contents for CONTURA
Page 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Page 62: ...CMM 4 8 61211 1310402 CONTURA...
Page 68: ...Installation 5 6 61211 1310402 CONTURA...
Page 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Page 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Page 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Page 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Page 130: ...Service features 8 8 61211 1310402 CONTURA...
Page 150: ...2Glossary 2 61211 1310402 CONTURA...
Page 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Page 157: ......
Page 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...