4.4
Operation Control of Spinning Disc Unit via AxioVision
Once electric power is available to the spinning disc unit, the Cell Observer SD can be completely
controlled via AxioVision software tools.
For information about AxioVision software control actions, you should consult the
AxioVision
Operating Manual
.
The shutter of the CSU-X1 spinning disc unit can also be released and closed by manual action. The
Shutter
key (Fig. 4-1/
1
) must be pressed for this purpose.
4.5
Switching Between Camera Ports
Spinning disc units with Dual Cam option allow you to switch between two camera ports.
•
Pull the push rod out (Fig. 4-1/
4
) to switch to the upper camera port.
If a secondary color splitter is installed, the short-wave emission light portion will be directed onto the
lateral camera adapter and the long-wave emission light portion onto the upper camera adapter when
the push rod is fully extracted.
•
Push the push rod in again, in order to work with the lateral camera port.
4.6
Exchange of Emission Filter
To exchange a currently installed emission filter,
proceed as follows:
•
Extract emission filter slide (Fig. 4-1/
5
) from the
spinning disc unit.
•
Use multi-purpose tool (Fig. 4-9/
1
) to unscrew
the retainer ring (Fig. 4-9/
2
). Take the filter out
3
).
•
Insert a new filter.
•
Screw retainer ring (Fig. 4-9/
2
) on again.
•
Introduce emission filter slide (Fig. 4-9/
4
) into
spinning disc unit until mechanical stop
position.
Fig. 4-9
Exchange of emission filter