2-29
2.2.14
Changing the filter set in the FL
P&C reflector module
The filter sets for the FL P&C reflector module can
be combined and assembled individually by the
customer. Appropriate filter sets or fully assembled
FL P&C reflector modules can be ordered from Carl
Zeiss.
•
Remove FL P&C reflector module (2-43/
3
) from
the reflector turret (also see 2.2.13).
•
Use mounting device from the tool set to
unscrew retainer ring (2-43/
).
•
Turn the reflector module around and allow the
filter (2-43/
or
4
) to drop on a soft surface.
•
The barrier filter is inserted at (2-43/
2
), the
4
), and both are
secured using retainer ring (2-43/
).
2.2.15
Changing the beam splitter in the
FL P&C reflector module
Attachment of filters and of the beam splitter requires utmost care to prevent damage and
contamination of the optical components.
We would recommend you to order completely equipped FL P&C reflector modules, since changing the
beam splitter requires much skill.
Otherwise, proceed as follows:
•
Remove FL P&C reflector module from the
reflector turret (also see 2.2.13).
•
Loosen the two slotted screws (2-44/
1
) with a
screwdriver.
•
Hold both halves of the reflector module
together turn opposite to the installation
position and place down.
Fig. 2-43
Changing the filter set in the P&C
FL reflector module
Fig. 2-44
Changing the beam splitter