OPERATION
Axioskop
2
plus
Carl Zeiss
Switching on and basic settings
Axioskop 2
mot plus
3-18
B 40-075 e 02/01
3.2
Switching on and basic settings
3.2.1 Axioskop
2
plus
•
Switch the instrument on with the on / off switch on the instrument rear (3-1/
27
).
The green pilot lamp (3-1/
) lights up.
•
Set the required brightness using the control for light intensity. The LED array (3-1/
) indicates the
present lamp voltage of the transmitted-light halogen illuminator.
3.2.2 Axioskop
2
mot plus
•
Switch the instrument on with the on / off switch on the instrument rear (3-1/
27
).
After the microscope is switched on, the control unit will perform a system check of all the connected
components. During the system check, the pilot lamp on the stand lights up in orange. After approx. 3
seconds, the microscope is ready for operation and is in the so-called automatic mode. The pilot lamp on
the stand now lights up in green.
The Axioskop 2
mot plus
features two different function modes and one programming mode:
Function modes:
−
In the
automatic mode
(standard setting), the illumination is set automatically (light manager
function) in accordance with the values last stored.
−
In the
manual mode
, the function of the light manager is switched off. Illumination intensity and
condenser settings must be made manually, as in the Axioskop 2
plus
.
Programming mode:
−
In the
programming mode
,
the user can individually match and store the focusing speeds and focus
positions of the various objectives (parfocality).