SET-UP
Carl Zeiss
Making electrical connections
Axio Zoom.V16
36
435080-9030-701
05/2012
3.12
Making electrical connections
Do not switch the microscope on until all cables have been connected.
Fig. 25 Connecting the microscope
•
Join the CAN bus connections of the CAN components used in any sequence to the
CAN A
to
CAN D
ports of the focusing drive.
In the absence of the motorized focusing drive (focus motor) each electronic CAN component
can be operated with a separate PSU. In this case, the body of the Axio Zoom.V16 microscope
must be connected directly to the mains via the plug-in power supply.
If two or more CAN components are connected to each other, only
one
power supply unit may
be used.
Item No.
Name/Order No.
Figure
Remarks
CAN cable RJ 45
Included with each
CAN component
Connection between focus
motor 3 and body of the
Axio Zoom.V16 microscope
and between focus motor 3
and MaRC
L=500 mm or
L=1000 mm
Plug-in power
supply
Power supply to focus motor
3 or body of the Axio
Zoom.V16 microscope with
coarse/fine drive