SET-UP
Axio Zoom.V16
Mounting transmitted light unit S or … (Option)
Carl Zeiss
05/2012
435080-9030-701
31
3.9
Mounting transmitted light unit S
•
Remove the base insert plate (Fig. 20/
3
). To do this loosen the short socket head screws (Fig. 20/
2
), lift
the insert plate (use recessed grip on the right edge of stand base (Fig. 20/
5
) for the purpose) and lift
securely with both hands.
•
Loosen the three socket head screws (SW 3) on the adapter ring (Fig. 20/
4
) using a ball-headed
screwdriver and remove the adapter ring.
•
Insert the transmitted light unit (Fig. 20/
6
) onto the stand base. The two large plastic taper pins on the
bottom of the transmitted light unit must click into the stand base.
•
Use the ball-head screwdriver (Fig. 20/
9
) to screw down the four countersunk socket head screws
(SW 3)
hand-tight on the transmitted-light unit.
•
Lay the base insert plate (Fig. 20/
3
) onto the transmitted light unit (Fig. 20/
6
).
•
Insert the opal glass plate
∅
84 mm (Fig. 20/
8
) to prevent screws and other small parts from being lost
in the course of further installation work.
•
Use the three short socket head screws (Fig. 20/
2
) to fasten the base insert plate (Fig. 20/
3
) (without
stage adapter) to the transmitted-light equipment.
If a stage is to be used, see Section 3.7, page 23.
•
Replace the opal glass plate
∅
84 mm (Fig. 20/
8
) with insert 52/84 (Fig. 20/
7
).
•
Insert glass plate 120 mm (Fig. 20/
1
); press on rear edge to remove it again.
Fig. 20 Mounting transmitted light unit S