SET-UP
Axio Zoom.V16
Mounting the stages
Carl Zeiss
05/2012
435080-9030-701
27
3.7.3
Stages with an 84 mm mounting
face
Stages with an 84 mm mounting
diameter must be attached to the
stand base with the aid of adapter
84/120 or 84+120 (see Section 3.7.2).
•
Insert the stage (Fig. 14/
2
) into stage adapter
4
) and align.
•
Throw eccentric clamping lever (Fig. 14/
3
) over
to secure the stage in the stage adapter.
•
Place the insert plate (Fig. 14/
1
) into the stage
interface.
3.7.4
Stages with an 120 mm mounting
face
On transillumination base 300 adapter
84+120 must be mounted for holding
stages (see Section 3.7.2). This is not
required for base 300 for reflected
light and stand base 450.
Securing with clamps
•
Insert stage with mounted clamping ring
(Fig. 15/
2
and
3
) into the adapter (Fig. 15/
5
) or
stand base and align.
•
Secure the clamping ring (Fig. 15/
3
) in the
adapter ring by turning the socket head screw
(SW 3) (Fig. 15/
4
) clockwise.
•
Install the insert frame or plate (Fig. 15/
1
).
Securing with screws
•
Set stage (Fig. 15/
6
) on the adapter or stand
base. Ensure that any existing orientation
elements on the underside of the stage engage
correctly in the corresponding grooves.
•
Secure the stage with the three captive screws
(Fig. 15/
7
).
•
Install the insert frame or plate (Fig. 15/
8
).
Fig. 14 Mounting stage with 84 mm mounting
face
Fig. 15 Mounting stage with 120 mm
mounting face on transillumination
base 300