OPERATION
Carl Zeiss
Operating the motorized microscope via the touchscreen of the TFT display
Axio Imager
102
430000-7344-001
M70-2-0020 e 06/2009
−
Load Position
button:
When this button is pressed, the stage travels to
the load position. The load position lies
approximately 10 mm below the operating
position. The load travel of the stage can be
interrupted via the
STOP
button (Fig. 4-25) or
via any button located on the stand base. When
the stage arrived in the load position, the
Load
Position
popup window (Fig. 4-26) appears
with the following control elements:
Stage returns to operating
position.
While the button is depressed,
the stage will travel to the
operating position (to the upper
stage stop at the most).
Stage moves down as long as
you press this button (to the
lower stage stop at the most).
When this button is pressed, the
current focus position is
adopted as the operating
position.
Caution! If the operating position set is
below the current focus position, the
focusing drive will move downward
when the
button is pressed.
When the
STOP
button is pressed, the load travel
is aborted and a pop-up window (Fig. 4-27) is
displayed. The additional button appearing here
serves to move the stage to the load position
originally intended.
When the lower stage stop is reached, the
message "Lower stage stop reached" will appear.
Fig. 4-25
STOP button
Fig. 4-26
Stage in load position
Fig. 4-27
Travel aborted