Axio Imager 2
FIRST-TIME SET-UP
ZEISS
Mechanical stage 75x50 mot. CAN
01/2016
430000-7544-001
79
3.37.2
Assembling mechanical stage 75 x 50 mot. CAN and connecting joystick / trackball
On delivery, the mechanical stage has a transport lock. This should be removed before attaching
the stage, as otherwise the mechanical stage could be damaged.
•
Unpack the mechanical stage and remove the transport lock (Fig. 70/
2
).
•
Place mechanical stage (Fig. 70/
1
) onto the stage carrier so that the holes in the underside of the
mechanical stage coincide with the through holes of the stage carrier.
•
Using the angled Allen key (AF 3), screw four screws (Fig. 70/
6
) from the bottom into the underside of
the stage, with the shorter screws being inserted in the front holes.
•
Remove cover (Fig. 72/
1
) by loosening the two screws.
•
Depending on which is used, insert the CAN connecting plug (Fig. 72/
5
) of the joystick (Fig. 72/
6
) or
the trackball (Fig. 72/
7
) into the connection socket on the mechanical stage (Fig. 72/
4
).
•
Connect the CAN connection socket on the mechanical stage (Fig. 72/
2
) to that on the rear of the
stand (Fig. 72/
8
) using a CAN-bus cable (Fig. 72/
3
).
•
Replace the cover (Fig. 72/
1
).
Fig. 72
Assembling mechanical stage 75 x 50 mot. CAN and connecting joystick / trackball