Axio Imager 2
FIRST-TIME SET-UP
ZEISS
Mounting the microLED attachment lamp for transmitted light
01/2016
430000-7544-001
65
3.31.2
Adjusting the VIS-LED attachment
lamp
•
Switch on the attachment lamp via the
microscope.
•
Switch off the diffusion disk for transmitted
light and remove the filter wheels for
transmitted light.
•
Use
≤
40x objective to focus on a free area of
the specimen.
•
Remove the eyepiece. Use the adjusting screws
(Fig. 51/
1
and
2
) to center the image of the
illuminator in the pupil image.
•
Use adjusting screw (Fig. 51/
3
) to achieve
optimally even illumination of the pupil image.
•
Switch the diffusion disk for transmitted light
on again and insert the filter wheels for
transmitted light.
3.32
Mounting the microLED
attachment lamp for transmitted
light
The microLED attachment lamp is only intended for
transmitted light operation.
Do not look directly into the LED light.
Only insert or remove the microLED
attachment lamp in/from the
transmitted-light port while switched
off.
•
Loosen clamping screw (Fig. 52/
1
) on
transmitted-light port (Fig. 52/
2
). Remove the
halogen lamp.
•
Use an AF 3 ball-headed screwdriver and
clamping screw (Fig. 52/
1
) to lock the
attachment lamp (Fig. 52/
3
) with dovetail
mount firmly into socket (Fig. 52/
2
).
•
Connect the three-pin illuminator plug (Fig. 52/
4
) to the lower three-pole 12 V / 100 W socket
5
) for transmitted light (TL) on the back of the instrument.
•
For the coded stand, set the transmitted-light / reflected-light toggle switch to transmitted light (TL). If
the motorized stand is used, switch between reflected light and transmitted light using the TFT display
touch screen.
The microLED must be configured as VIS-LED on the Axio Imager .M2 / .Z2.
Fig. 51
Adjusting the VIS-LED attachment
lamp
Fig. 52
Mounting the microLED attachment
lamp