CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axio Imager 2
Instrument maintenance
198
430000-7544-001
01/2016
5.2.3
Changing fuses on the VP232-2
CAN power supply
Always disconnect the power plug
before changing the fuses.
If any fuses fail, the cause of the failure must first
be determined. Any technical defect should be
repaired by trained personnel.
The fuse compartment is located on the back of
the power supply and contains two type
T 4.0 A /
250 V, 5x20 mm fuses
.
•
Disconnect the power plug.
•
Pull out fuse holder (Fig. 214/
2
) forward. Use a
small screwdriver, if necessary.
•
Remove the fuses from the holder and insert
new fuses.
•
Push the fuse holder back into the fuse
compartment (Fig. 214/
1
) as far as it will go.
•
Connect the power plug.
Fig. 214
Changing fuses on the power
supply