MAINTENANCE, REPLACEMENT OF FUSES AND SERVICE
5.1
Maintenance
Before cleaning, switch off the microscope and the fc12 Auto Focus controller and disconnect
the external power unit 232-2 and the Auto Focus controller from the power supply.
The controller and the Auto Focus sensor should only be cleaned with a slightly moist cloth (not dripping
wet). No moisture should enter the devices.
Only clean the type plate and the warning sticker with pure water.
5.2
Replacement of fuses in the fc12
Auto Focus controller
Make sure the power cable is
disconnected before changing any
fuses.
In the event of a fuse failure, the cause must be
identified and a possible technical fault must be
first remedied in a proper manner.
The fuse holder is located at the back of the
controller. It is integrated into the power plug and
contains two
T 2.0 A/H
fuses.
•
Pull the mains plug.
•
Remove the fuse holder (Fig. 15/
1
) from the
fuse box by pulling it towards you. A small
screwdriver is suggested for this purpose.
•
Remove the fuses from the fuse carrier and
replace them with new fuses (type
T 2.0 A/H
).
•
Push the fuse holder back into the fuse box as
far as it will go.
•
Reconnect the mains plug.
1
Fuse carrier
Fig. 15 Replace the fuses in the controller