The
FIB
tab of the
FIB Control
panel opens.
IMPORTANT
Do not change the FIB Extr. Target value.
Changing this value would require a complete adjustment of the FIB probe currents.
5
Ensure the
Regulate
checkbox is activated.
This guarantees a stable emission current.
The emission is automatically regulated by chang-
ing the
FIB Suppressor Target
which can have val-
ues between -2000 V and 0 V.