pbj=çéÉê~íáçå
6.4.2. Setting detection parameters
6.4.2.1. Selecting a detector
1
Go to the
Detectors
tab of the
SEM Control
panel.
2
Select the detector from the
Detectors
drop-
down menu.
The following table should serve as a help to find the required settings for your application.
Detectors
EHT
Typical
WD
Detector settings
Remarks
In-lens
3 kV - 20 kV
3 - 6 mm
None
100 V - 3 kV
2 - 3 mm
100 V
max. 4 mm
SE2
1 - 30 kV
min. 4 mm
Collector voltage (= bias) adjustable
from -250 V to + 400 V
Standard applications: +300 V
Pseudo BSE image: -150 to 0 V
Select the settings as
described in section 6.4.2.2.
1 kV - 5 kV
4 - 6 mm
5 kV - 30 kV
min. 6 mm