25
Changing the motor
AURA
®
handheld NIR - Rev. 2
EN_40_070_0030I | AURA
®
handheld NIR - Rev. 2 | Maintenance and Repair Manual
11
Lay cable and insert front panel
Pull the front panel
4
slightly out of the housing.
Plug the flat control cable
4.2
into slot
X704
4.1
on the “FBG PowerBoard”.
Lay the motor cable
3.2
so that it runs below the lamp cable along the
bottom of the housing.
Note
NOTE ON LAYING THE MOTOR CABLE
It is much easier to lay the cables if the lamp holder is temporarily
dismantled / removed from the housing.
At the narrow points (see red dotted ellipse in Fig. 23 on page 26) , take
special care to avoid damaging the electronic components or cables.
Plug the motor cable
3.2
into slot
X800
3.1
.
Insert the front panel
2
very carefully into the housing from above
1
.
Ensure that the flat control cable is laid so that it runs through the two
recesses on the underside of the sheet metal
2.1
of the front panel
2
(see the two red dashed circles in in Fig. 23 on page 26).
Fig. 22
X800
X600
X704
X702*
2.1
2
1
3.1
4.1
3.2
4.2
Lay the cable and insert the front panel