23
Changing the motor
AURA
®
handheld NIR - Rev. 2
EN_40_070_0030I | AURA
®
handheld NIR - Rev. 2 | Maintenance and Repair Manual
8
Install new motor
Insert the motor into the motor holder
2
right up to the mechanical stop
1
.
Turn the motor so that the screw
5
of the semicular mount faces upwards
and the cables
4
of the motor point downwards.
Screw the two clamping screws
3
relatively loosely into the motor holder
so that the motor can still be moved.
Fig. 19
3
2
1
5
4
Install new motor
9
Attach the white reference assembly
Place the white reference assembly
2
onto the white reference mount of
the motor shaft
1
.
Secure the white reference assembly
2
using the two Torx screws
3
.
Use a TX 8 Torx screwdriver for this.
Fig. 20
1
2
3
Attach the white reference assembly