4 Installation
Instruction for use
AT 030
4.5 Mounting the device on SL 120, SL 130, SL 220, and ZEISS VISULAS laser slit
lamps using the fixed tonometer holder
20 / 60
319300-9120-000-GA-GB-240320
4.5 Mounting the device on SL 120, SL 130, SL 220,
and ZEISS VISULAS laser slit lamps using the
fixed tonometer holder
NOTE
Damage due to improper assembly
u
After assembly, make sure the device and tonometer holder are
firmly seated.
Action
1. Remove the cap from the swivel arm of the slit projector.
2. Plug the centering pivot of the tonometer holder into centering
aperture so that mounting plate faces the patient side.
3. Attach the applanation tonometer to the base plate of the
tonometer holder and fix it by means of the knurled screw.
4. Ensure that the centering pin is securely seated into one of the
two alignment holes. Select the alignment hole on the side
where the eyepiece is to be used.
NOTE! The measuring prism must be visible in the center of
the eyepiece being used. Correct any lateral misalignment of
the tonometer by slightly turning the tonometer holder.
If the correct position of the tonometer has been found, fasten the
tonometer holder with the two fastening screws in the holes
provided.
HINT: If the device is used on ZEISS slit lamps using the fixed
tonometer holder it cannot be swung out of the observation path.
The device may, however, be easily removed from the slit lamp
prior to diagnosis or treatment by loosening the knurled screw. The
tonometer holder can remain on the slit lamp.
Summary of Contents for AT 030
Page 1: ...AT 030 Applanation tonometer Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...AT 030 Applanation tonometer Instruction for use...
Page 8: ......
Page 16: ...Empty page for your notes...
Page 20: ...Empty page for your notes...
Page 28: ...Empty page for your notes...
Page 36: ...Empty page for your notes...
Page 38: ...Empty page for your notes...
Page 52: ...Empty page for your notes...
Page 56: ...Empty page for your notes...
Page 60: ...Empty page for your notes...
Page 62: ...Empty page for your notes...
Page 64: ...Empty page for your notes...
Page 68: ......
Page 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Page 70: ......
Page 72: ...Empty page for your notes...
Page 85: ...Empty page for your notes...
Page 87: ...Empty page for your notes...