4 Installation
Instruction for use
AT 030
4.4 Mounting the device to the SL 120, SL 130 and SL 220 slit lamps using the
swivel-mounted tonometer holder behind the stereomicroscope
18 / 60
319300-9120-000-GA-GB-240320
4.4 Mounting the device to the SL 120, SL 130 and
SL 220 slit lamps using the swivel-mounted
tonometer holder behind the stereomicroscope
NOTE
Damage due to improper assembly
u
After assembly, make sure the device and tonometer holder are
firmly seated.
NOTE! If accessory units have been mounted between the
microscope and tube, the tonometer holder must always be
attached directly to the microscope!
Action
1. Remove the binocular tube from the slit lamp.
2. Attach the tonometer holder to the slit lamp.
3. Attach the applanation tonometer to the base plate of the
tonometer holder and fix it by means of the knurled screw.
4. Ensure that the centering pin is securely seated into one of the
two alignment holes. Select the alignment hole on the side
where the eyepiece is to be used. If the tonometer holder is
adjusted correctly the measuring prism will be visible in the
center of the eyepiece in use.
HINT: The applanation tonometer with the base plate of the swivel-
mounted tonometer holder can easily be swung out of the optical
path by turning it laterally to the right around its vertical axis A.
This enables unhindered work with the slit lamp. Additionally, the
extensible measuring prism holder can be pushed into the
tonometer. The measuring prism holder should then be in its
bottom end position.
Summary of Contents for AT 030
Page 1: ...AT 030 Applanation tonometer Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...AT 030 Applanation tonometer Instruction for use...
Page 8: ......
Page 16: ...Empty page for your notes...
Page 20: ...Empty page for your notes...
Page 28: ...Empty page for your notes...
Page 36: ...Empty page for your notes...
Page 38: ...Empty page for your notes...
Page 52: ...Empty page for your notes...
Page 56: ...Empty page for your notes...
Page 60: ...Empty page for your notes...
Page 62: ...Empty page for your notes...
Page 64: ...Empty page for your notes...
Page 68: ......
Page 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Page 70: ......
Page 72: ...Empty page for your notes...
Page 85: ...Empty page for your notes...
Page 87: ...Empty page for your notes...