ACCURA Installation site requirements
25
ACCURA 20/24/10 installation dimensions
1064
1151
660
2367
3067
3140
3610
640
2250
2760
2460
≥
200
÷
All dimensions in millimeters, scale 1:50
600
680
≥
800
≥
800
CMM width
Length of granite plate, approx. 130 mm extra for
service unit.
CMM height
Height of secured CMM assembly with shipping pal-
let.
Height of secured CMM assembly from lower edge of
granite plate.
Height between the granite plate and lower edge of
the stylus mounting cube on the VAST XT gold
probe.
With other VAST gold or RDS probe types, the height
is somewhat less.
Network connection for data backup and Internet
(Teleservice)
Power supply
Compressed air supply
TCC (standard) controller cabinet, MCC 800 (option)
➤
see Controller cabinet installation [
CMM <-> controller connection cable (cable length
approx. 3 m)
Compressed air connection on the CMM
Scanner protective field option
Weight:
ACCURA (model 20/24/10)
13000 kg
Workpiece, max.
5000 kg