background image

ACCURA Installation site requirements

23

ACCURA 16/42/15 installation dimensions

4940

2627

2020

1927

670

1651

1564

640

4616

200

3260

2750



‘

Œ



Ž

÷

All dimensions in millimeters, scale 1:50

600

680

800

†

‚



„

…

‡

ƒ

800

Œ

CMM width



Length of granite plate, approx. 130 mm extra for service
unit.

Ž

CMM height



Height of secured CMM assembly with shipping pallet.



Height of secured CMM assembly from lower edge of granite
plate.

‘

Height between the granite plate and lower edge of the sty-
lus mounting cube on the VAST XT gold probe.

With other VAST gold or RDS probe types, the height is
somewhat less.



Network connection for data backup and Internet (Teleser-
vice)

‚

Power supply

ƒ

Compressed air supply

„

TCC (standard) controller cabinet, MCC 800 (option) 

 

see

Controller cabinet installation [

 18]

…

CMM <-> controller connection cable (cable length approx. 3
m)

†

Compressed air connection on the CMM

‡

Scanner protective field option

Weight:

ACCURA (model 16/42/15)

18500 kg

Workpiece, max.

4000 kg

Summary of Contents for ACCURA

Page 1: ...ACCURA Installation site requirements Coordinate measuring machine CMM Size 16 24 15 16 30 15 16 42 15 20 24 10 20 30 10 20 42 10 20 24 15 20 30 15 20 42 15...

Page 2: ...ACCURA Installation site requirements 2021 10 11 61221 1041202...

Page 3: ...alled CMM 10 Planning and Measuring Lab Preparation 11 General protection measures 11 Doorway dimensions in the measuring lab 11 Room height of the measuring lab 11 Floor 11 Floor plan 11 Vibrations a...

Page 4: ...20 42 10 bellows cover installation dimensions 30 ACCURA 20 24 15 installation dimensions 31 ACCURA 20 24 15 bellows cover installation dimensions 32 ACCURA 20 30 15 installation dimensions 33 ACCURA...

Page 5: ...ments 5 Limit curves of permissible floor vibrations at the installation site 47 Permissible foundation acceleration for ACCURA X1600 with active damping 47 Permissible foundation acceleration for ACC...

Page 6: ...ACCURA Installation site requirements 6...

Page 7: ...Coordinate measuring machine CMM Controller cabinet Control console Data system and peripheral devices optional Note on data systems NOTE Data systems for coordinate measuring machines are CMM based s...

Page 8: ...unload and transport to the storage location or installation site suitable transportation equipment and operat ing personnel must be provided The type of transport equipment depends on the local cond...

Page 9: ...0700 Truck Pallet 5680 2900 3250 1240 19740 20 24 10 Air sea Crate 3870 3180 2970 36 5 1665 14365 Truck Pallet 3870 3180 2800 960 13660 20 30 10 Air sea Crate 4500 3180 2970 42 5 2240 17240 Truck Pall...

Page 10: ...only be removed by a ZEISS service engineer NOTE Packing materials such as disposable transport packaging for the covers PC and accessories foil or Styrofoam chips can be returned to ZEISS if they are...

Page 11: ...a floor plan ensure that your coordinate measuring machine can be easily accessed from all sides Installation near transport routes must be avoided Vibrations at the installation site Vibrations are...

Page 12: ...on Note Humidity values that exceed the permissible range require an additional air conditioner on the con troller cabinet 40 70 Ambient temperature 10 35 C Height above mean sea level 100 125 V 3000...

Page 13: ...10 l min at 5 bar 50 Nl min at 1 bar Compressed air connection point The compressed air line must be equipped with a lockable separator and a pressure release system for the compressed air connection...

Page 14: ...against being turned on again uninten tionally Warning some components in the controller cabinet remain live even when the main switch is turned off These places are marked with warning signs If nece...

Page 15: ...iving areas In such cases opera tors may be required to implement appropriate measures at their own expense For peripheral devices e g computer monitor etc the specifications of the respective manufac...

Page 16: ...connection with a circuit breaker Alternatively equivalent protective equipment can also be used We rec ommend installing a lightning arrester e g Phoenix Flashtrab FLT 25 400 for general protection...

Page 17: ...current circuit breaker 4 Cable included with delivery 5 Main switch 6 Shock proof outlet Note For maintenance work on the CMM Network connection A network connection is required for Data backup Teles...

Page 18: ...cted towards the CMM Only for MCC 800 the distance between the back of the controller cabinet and the wall must be at least 500 mm When the back is fully open the distance to the wall must be at least...

Page 19: ...lower edge of granite plate Height between the granite plate and lower edge of the sty lus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the height is somewhat less N...

Page 20: ...ecured CMM assembly from lower edge of granite plate Height between the granite plate and lower edge of the sty lus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the...

Page 21: ...lower edge of granite plate Height between the granite plate and lower edge of the sty lus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the height is somewhat less N...

Page 22: ...ecured CMM assembly from lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the h...

Page 23: ...lower edge of granite plate Height between the granite plate and lower edge of the sty lus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the height is somewhat less N...

Page 24: ...ecured CMM assembly from lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the h...

Page 25: ...lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the height is somewhat less N...

Page 26: ...secured CMM assembly from lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the...

Page 27: ...lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the height is somewhat less N...

Page 28: ...secured CMM assembly from lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the...

Page 29: ...lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the height is somewhat less N...

Page 30: ...secured CMM assembly from lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the...

Page 31: ...lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the height is somewhat less N...

Page 32: ...secured CMM assembly from lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the...

Page 33: ...lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the height is somewhat less N...

Page 34: ...secured CMM assembly from lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the...

Page 35: ...lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the height is somewhat less N...

Page 36: ...secured CMM assembly from lower edge of granite plate Height between the granite plate and lower edge of the stylus mounting cube on the VAST XT gold probe With other VAST gold or RDS probe types the...

Page 37: ...tored A risk due to travel movements is thus reduced to a minimum Once the protected area is cleared the CMM moves again at maximum speed Using floor markings mark the boundary of the protective field...

Page 38: ...d information ACCURA with active damping Alignment of air dampers for X1600 The base is comprised of four air dampers and four leveling units which are suitable for screed floors as per DIN 18560 The...

Page 39: ...ce for center to center distance of the air dampers 10 mm Note The tolerances must not be exceeded Floor load Measuring range X Y Z mm Value N mm2 Mean surface pressure on the 4 support surfaces 1600...

Page 40: ...50 850 2890 750 0 0 2250 1100 3840 750 0 0 2250 530 530 384 3740 4940 3140 1 Outer edge of granite table 2 Air damper BiAir3 with leveling unit 3 Air damper BiAir4 with leveling unit Tolerances for th...

Page 41: ...70 1870 2070 2270 2470 0 Y X 474 0 1 2 3 4 5 6 7 8 148 5 200 1600 374 2470 2400 176 7 3140 M12 96x 2 1 3 674 874 1074 1274 1474 1674 1874 Left Y guideway Right Y guideway Measuring range Threaded inse...

Page 42: ...0 2870 0 Y X 474 0 1 2 3 4 5 6 7 8 148 5 200 1600 374 2470 3000 176 7 3740 M12 120x 2 1 3 674 874 1074 1274 1474 1674 1874 Left Y guideway Right Y guideway Measuring range Threaded insert 4 x M36 for...

Page 43: ...0 Y X 474 0 1 2 3 4 5 6 7 8 148 5 200 1600 374 2470 4200 176 7 4940 M12 168x 19 18 17 20 21 2 1 3 674 874 1074 1274 1474 1674 1874 Left Y guideway Right Y guideway Measuring range Threaded insert 4 x...

Page 44: ...5 705 905 1105 1305 1505 1705 1905 2105 2305 0 1 2 3 4 5 6 7 8 9 10 148 5 200 2000 400 2910 2400 176 7 3140 M12 120x 2 1 3 Left Y guideway Right Y guideway Measuring range Threaded insert 4 x M36 for...

Page 45: ...0 Y X 505 705 905 1105 1305 1505 1705 1905 2105 2305 0 1 2 3 4 5 6 7 8 9 10 148 5 200 2000 400 2910 3000 176 7 3740 M12 150x 2 1 3 Left Y guideway Right Y guideway Measuring range Threaded insert 4 x...

Page 46: ...0 505 705 905 1105 1305 1505 1705 1905 2105 2305 0 1 2 3 4 5 6 7 8 9 10 4200 176 7 4940 M12 210x 148 5 200 2000 400 2910 Y X 2 1 3 Left Y guideway Right Y guideway Measuring range Threaded insert 4 x...

Page 47: ...re an additional foundation 0 1 1 10 100 1000 0 10 20 30 40 50 Vibration acceleration mm s Frequency Hz X axis horizontal Y axis horizontal Z axis vertical Permissible foundation acceleration for ACCU...

Page 48: ...ACCURA Installation site requirements 2021 10 11 61221 1041202...

Reviews: