What to do in the event of malfunctions
S100 / OPMI pico
Version 7.0
Page 216
G-30-1781-en
Malfunctions of S100 suspension systems with video and image storage op-
tion
Fault
(Message/acoustic signal)
Possible cause
Remedy
Cross-
reference
No stick available
No USB data carrier inserted. Insert USB data carrier.
–
Not enough memory avail-
able
USB data carrier is inserted
but there is insufficient
memory space.
Press the Index key; menu ap-
pears on the screen, permit-
ting images to be deleted.
–
Write-protection activated
USB data carrier is inserted
but write-protection is active.
Deactivate write protection.
–
Defective USB memory
please remove the stick
USB data carrier does not
contain any file system or a
damaged
file system.
Replace the USB data carrier. –
Error
Remote Short Circuit -
Disconnect Input
Short circuit on Remote
socket - all keys have been
disabled.
Contact service.
–
The messages are automatically deleted.
Summary of Contents for 000000-1080-349
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...