background image

COMMISSIONING AND OPERATION (Fig. 3)

The green "Power on" light on the front panel of the unit indicates voltage is present while a yellow "Pump on" lamp indicates that the 

pump is operating. When the unit is plugged in, the green and yellow lights will turn on. the yellow light indicates that the pump has 

started (Fig. 3 / A); it remains in operation for a few seconds to pressurize the system. If this time is not sufficient, the red light comes 

on,  in  this  case  hold  down  the  red  button  Restart  and  wait  until  water  comes  out  of  an  opened  sampling  point.  After  closing  the 

sampling point, the device switches off the pump and then remains in standby mode; the green light stays on and the device is ready 

to carry out all further command and control steps in complete autonomy (Fig. 3 / B). When opening a sampling point, the device 

switches on the pump; This remains in operation until the sampling point is closed again (Fig. 3 / A). After closing the sampling point, 

the device restores the maximum pressure in the system, switches off the pump and returns to standby mode (Fig. 3 / B). If there is a 

lack of water during priming, the unit switches off the pump, signals it with the red light (Fig. 3 / C) and thus protects it from running 

dry.  Once  the  fault  that  caused  the  blockage  is  eliminated,  it  is  sufficient  to  press  the  red  "Restart"  button  to  resume  normal 

operation. In the case of a temporary interruption of the power supply, the device restarts automatically as soon as the power returns.

PROBLEMS

 

The pump does not start.

The pump does not turn on.

The pump runs intermittently.

The pump does not turn off.

The pump is blocked.

MAIN CAUSE

Faulty power connection.

The water column is too high.

Water loss of the system less than the minimum flow.

Water loss in the system greater than the minimum flow.

Priming problems.

PICTURES

Attention!

Pressure switch may only 

be mounted vertically 

(pointing upwards)!

Reviews: