For “

Normal

” mode, The oscilloscope will not refresh the display until there is a 

waveform trigger. Therefore, in this mode, the screen waveform may not refresh or 
there  is  no  waveform.  It  is  suitable  for  measuring  discontinuous  waveforms  and 
requires users to adjust the trigger level in advance. 

Trigger  type

  includes  rising  and  falling  edges.  Rising  edge  trigger  means  that 

the waveform crosses the trigger level line upward, and a trigger condition occurs. 

 

Trigger level

 include "automatic" and "manual". 

"Auto" is suitable for trigger mode = auto, the oscilloscope detects the waveform 

in real time and automatically adjusts the trigger level. 

"Manual"  When  the  “single”  is  used  or  the  trigger  mode  is  "Normal",  the 

"Manual" option should be used. Under this option, the trigger level can be manually 
adjusted by the user.

 

Trigger  source

 

indicates  which  channel's  waveform  is  currently  used  for 

triggering. In the dual-channel mode, CH1 or CH2 can be selected. If the channel is 
used as the trigger source, its waveform can be stable. If the waveform frequencies of 
the two channels are different, the waveform of the other channel will be unstable.

 

8.4

 

Display 

Persistence

”  is  to  make  the  historical  waveform  remain  on  the  screen  for  a 

while and then disappear. Optional persistence options are: None, 1s, Infinite. When 
Infinity  is  selected,  the  waveforms  that  have  appeared  on  the  screen  will  remain 
forever and will not disappear. 

roll

”  is  to  display  the  waveform  while  sampling.  Only  when  the  time  base  is 

greater than or equal to 200ms will the roll mode be entered. If roll is turned off, the 
waveform  will  be  displayed  after  one  screen  is  sampled,  resulting  in  a  long  time 
before the screen is refreshed for large time base. 

“XY”

  mode  is  a  way  that  channel  1  is  used  as  the  X  axis  and  channel  2  is 

displayed on the Y axis. It can be used to view 

Lissajous

 graphics etc. 

"

Hi-rate

" refers to the waveform refresh speed. When turned on, many frame 

waveforms will be superimposed. When turned off, only one waveform is displayed 
at a time. Note that it is always “on” when the time base <=200ns. 

8.5

 

Settings 

When the FFT option is turned on, the CH1 waveform can be FFT calculated and 

displayed.  FFT  display  options  are:  logarithmic,  linear,  and  music  spectrum. 
Measuring FFT results is not supported. 

The  calibration  option  is  used  to  calibrate  the  zero  point  of  the  oscilloscope. 

Self-calibration is required when the outside temperature changes or when there is a 
zero offset due to aging. Remove the probes before calibration. 

Furthermore: 
1. In the settings, you can set whether to automatically shut down, if you choose 

Summary of Contents for DSO3D12

Page 1: ...DSO3D12 User s Manual Designed by ZEEWEII DIGITAL OSCILLOSCOPE...

Page 2: ...de is normal Trigger edge Rising edge trigger or falling edge trigger Battery level indication Trigger level position X1 X10 indicates probe attenuation which should be set to match the probe The goal...

Page 3: ...eform 50 Quick centering CH1 DMM Turn the multimeter on off Press Shift DMM to full screen Select Multimeter function selection mV V Adjust vertical sensitivity voltage of one grid s ns Adjust the tim...

Page 4: ...connected the oscilloscope automatically adjusts the range in real time Press any key to exit Shift V ch1 Turn on off the voice assistant 1 Safety Precautions High voltage The x10 range of the probe...

Page 5: ...anual adjustment mV V buttons scale the waveform vertically The s ns button scales the waveform horizontally The vertical sensitivity indicates the voltage of a grid in the vertical direction and it i...

Page 6: ...set the high priority of channel 2 pay attention to the change of the 1 2 icon in the upper left corner When not using dual channel measurements Channel 2 should be turned off for higher sample rates...

Page 7: ...s 3 4 1 Large input capacitance Because the probe 1X has a large input capacitance usually more than 100 pf it is not suitable for measuring some capacitively sensitive circuits For example when measu...

Page 8: ...f the signal by observing how many divisions the waveform occupies on the screen For example if the vertical sensitivity is 1v div and the signal occupies 3 grids on the screen you can know that the a...

Page 9: ...for adjusting the time base in the oscilloscope button panel are ns and s button When observing high frequency signals you should press the ns button to decrease the time base When observing low freq...

Page 10: ...The arrow circled above is the trigger level position and when the edge of the square wave in the above figure passes upwards through the position of this arrow a trigger is formed the waveform only...

Page 11: ...ancel a single trigger click Stop to stop Procedure 1 Connect the signal first 2 mV V set the voltage sensitivity s ns set the time base 3 Set the trigger level to manual and adjust the trigger level...

Page 12: ...exit can adjust the duty cycle other waveforms do not work When the set frequency is greater than 1MHz the square wave signal may show small horizontal jitter Single dual channel button to turn on off...

Page 13: ...e 1X 8 2 Measurement There are 14 measurement options available as shown below Frequency Peak to Peak Duty cycle Amplitude RMS Average Period Pulse width Pulse width Max Min Top Base Duty cycle The di...

Page 14: ...emain on the screen for a while and then disappear Optional persistence options are None 1s Infinite When Infinity is selected the waveforms that have appeared on the screen will remain forever and wi...

Page 15: ...Support Equivalent sample 500M Screenshot Support Rise time 3ns Frequency 0 01 Storage depth 64K Single Support Impedance 1M FFT Support Time base 5ns 10s Wave out amp 2 5V 0 05 Ver sensitivity 10mV...

Page 16: ...ice speech rate accent etc we cannot guarantee the accuracy of recognition Wake up command Hello ZEEWEII Pronounced Hello see wave or Hello mini scope Command words function Auto setting It is equival...

Page 17: ...w do I reset the device to its initial state Answer Open Menu Set and then long press the Auto button to reset 1 What is the difference between probe 1X and 10X Answer The 1X does not attenuate the si...

Page 18: ...cted to the audio output The time base is adjusted to about 1 10ms and the voltage sensitivity is adjusted to about 500mV decided according to the volume 6 How to measure main board communication sign...

Reviews: