background image

 

14 

 

The  difference  between  peak-to-peak  and  amplitude(AMP)  is  shown  in  the 

figure above. 

7.3  Trigger 

There  are Three options in the trigger menu: “Trigger Mode” /  “Trigger Type” 

and “Trigger level”. 

Trigger Mode

 can be selected from "Auto" and "Normal". 

For “Auto” mode, if a waveform trigger is detected, the waveform signal will be 

used  for  triggering.  If  the  oscilloscope  cannot  detect  the  trigger  condition,  it  will 
automatically force the trigger to display the waveform. 

For  “Normal”  mode,  the  oscilloscope  will  not  refresh  the  display  until  a 

waveform trigger occurs. So there may be no waveform on the screen in this mode. 

Trigger  type

  includes  rising  and  falling  edges.  Rising  edge  trigger  means  that 

the waveform crosses the trigger level line upward, and a trigger condition occurs. 

 

Trigger level

 include "automatic" and "manual". 

"Automatic"  is  suitable  for  novices.  The  oscilloscope  detects  the  waveform 

position and amplitude and automatically sets the trigger level. 

"Manual"  is  suitable  for  people  who  are  familiar  with  the  oscilloscope.  When 

using  the  “Single”  function  or  the  trigger  mode  is  "Normal",  you  should  use  the 
"manual" option. Under this option, the trigger level can be manually adjusted by the 
user. 

7.4  Display 

In this menu you can set persistence / roll mode and brightness. 

Persistence

”  is  to  make  the  historical  waveform  remain  on  the  screen  for  a 

while and then disappear. Optional persistence options are: None, 1s, Infinite. When 
Infinity  is  selected,  the  waveforms  that  have  appeared  on  the  screen  will  remain 
forever and will not disappear. 

roll

”  is  to  display  the  waveform  while  sampling.  Only  when  the  time  base  is 

greater than or equal to 200ms will the roll mode be entered. If roll is turned off, the 
waveform will be displayed after one screen is sampled at any time base, resulting in 
a long time before the screen is refreshed for large time base. 

7.5  Aux 

There are two options of "Calibration" and "FFT" in the auxiliary menu. 

When  the  probe  is  removed,  if  there  is  a  significant  deviation  between  the 

waveform position and the zero arrow position, it indicates that calibration is required. 

Please  remove  the  probe  and  USB  cable  before  calibration,  and  then  click  the 

"OK"  button.  The  oscilloscope  then  starts  self-calibration.  Do  not  perform  any 
operation on the oscilloscope until it is completed. After completing the calibration, 

Summary of Contents for DSO1511G

Page 1: ...ZEEWEII DSO1511G User s Manual Designed by zeeweii DSO1511G HANDHELD OSCILLOSCOPE...

Page 2: ...enerator When it is closed no display Time base which means the time of a grid in the horizontal direction Click the ns and s buttons to adjust Coupling DC AC Probe attenuation x1 x10 x100 Vertical se...

Page 3: ...r buttons Power OK short press Press and hold Power and short press Ok to switch the trigger edge rising or falling Power OK long press Press and hold Power and long press Ok to switch the trigger mod...

Page 4: ...VITY 7 3 2 VERTICAL POSITION 8 4 HORIZONTAL SYSTEM 8 4 1 TIME BASE 8 4 2 HORIZONTALLY MOVING WAVEFORM 8 5 TRIGGER SYSTEM 9 5 1 TRIGGER LEVEL 9 5 2 TRIGGER POSITION 9 6 OPERATION GUIDE 10 6 1 POWER 10...

Page 5: ...danger be sure to use this product as specified Only authorized personnel should perform maintenance procedures Avoid fire and personal injury Use the probe correctly and ensure that the measured vol...

Page 6: ...thout overshoot 3 Adjust the horizontal and vertical scales of the oscilloscope so that the waveform is displayed stably at a suitable position on the screen 4 Use a screwdriver to rotate the adjustab...

Page 7: ...gh voltage signal otherwise the high voltage may break down the internal circuit of the oscilloscope 2 4 3 High bandwidth Because the 1X probe has a large input capacitance the bandwidth of the probe...

Page 8: ...observing how many grids a waveform period occupies For example the current time base is 500us and one cycle of the signal occupies 2 grids so the signal period is 1ms 1KHz The oscilloscope time base...

Page 9: ...In edge trigger mode a trigger condition occurs only when the waveform crosses the trigger level To adjust the trigger level 1 Change Trigger Level to Manual in the menu 2 Method 1 Press and hold the...

Page 10: ...rging current is about 700ma The LED at the bottom of the charging process is bright red and the led turns green when fully charged 6 3 Vertical Time base scale Press the mV and V buttons to adjust th...

Page 11: ...s sampling is not completed until there is a waveform trigger Operation method Click the xKey button and select Single 6 9 Reference waveform The reference waveform is that the waveform of the previou...

Page 12: ...square wave signal may show small horizontal jitter 6 13 Video output The 3 5mm headphone jack at the bottom can output AV analog video signal which can be connected to a TV or projector Open Menu Set...

Page 13: ...ltage of the dotted line below dY Y1 Y2 is the voltage difference between the two dotted lines 7 Menu introduction Click the Menu button to open close the menu 7 1 Channel The channel menu includes tw...

Page 14: ...rigger mode is Normal you should use the manual option Under this option the trigger level can be manually adjusted by the user 7 4 Display In this menu you can set persistence roll mode and brightnes...

Page 15: ...after no button operation 2 You can choose whether to turn on the sound 3 Language selection Chinese or English 4 In this menu long press the ns button to enter the usb connection state long press s...

Page 16: ...16...

Reviews: