YZ Systems Milton Roy • 201 Ivyland Road • Ivyland, Pennsylvania • USA • 18974 • P: 281.362.6500 • www.yzsystems.com
DP-2010 ver.05-2018
Page 12
s
eCTion
6: P
rogramming
f
or
P
roPorTional
-T
o
-
T
ime
o
PeraTion
1. Calculate the sampling rate to fill a cylinder in 30 days:
2. Set the timer dials on the Z-65 to the sample rate
from step 1.
NOTE:
To obtain maximum battery life, choose
the longest time interval and the largest pump
displacement setting possible.
NOTE:
The time (18 minutes), in the chart
above, corresponds to the dial setting shown for
the Z-65 model with the timer range setting in
the factory position (jumper on the two left pins).
See section 10 Timer Range Setting, Page 24
.
figure 9
Example
Number of
turns
open on pump
stroke knob
Sample
pump
displacement
per stroke
Sample cylinder volumes
1000 cc
500 cc
300 cc
3
.100
4
9
15
6
.200
9
18
30
9
.300
13
27
45
12
.400
18
36
60
Sample
rate
(minutes)
3. Adjust the pump volume adjustment knob to the value used in the calculations in step 1.
Sample pump displacement per stroke
Number of turns open on the pump volume knob
.1cc
3
.2cc
6
.4cc
12
4. Turn mode switches 1, 2, 3 to ON and 4 to
OFF.
5. Press the test button once to initiate the
timer sequence.
figure 10