Yuwell YH-550 User Manual Download Page 9

13

14

and 3 is the highest level of FPS. The higher FPS level means more comfortable expiration you will 
get.

 

y

Date

: Adjust the date (year, month, day) displayed on the device.

 

y

Time

: Adjust the time (hour, minute) displayed on the device.

 

y

Language

: You can choose English or Chinese.

 

y

Filter Cotton Reminder

: When this function is enabled, this can remind you to check and change 

the filter.

 

y

Reset

: When this function is enabled, parameters of the device will be reset to the factory defaults 

and data will be erased.

Info

Humidity

Setup

Ramp

Sound Reminder
Smart Start/Stop

OFF

FPS Level

0

Date

2019-5-4

OFF

         

                                    Picture 5-1: Home                                  Picture 5-2: Setup

Date

2019-5-4

Time

9:27

Language

EN

OFF

Filter Cotton Reminder

Language

EN

Filter Cotton Reminder OFF
Reset
Back<<

>>

         

                                    Picture5-3: Setup                                     Picture 5-4: Setup

 

f

11.2  Function that patient must use under the direction of doctor (Clinical menu)

The patient should use the device by the direction of physician.

① Set Mode (CPAP or APAP)

 

y

At home page, press 

 and dial at the same time, the clinical menu will be shown on screen.

 

y

Rotate the dial, highlight the "Mode" , and then select CPAP or APAP.

 

y

After mode setting, press dial to save your change.

 

y

Highlight the "Back", press dial to back to home page.

② Set Pressure 

 

y

At clinical page, choose "CPAP" mode, Highlight "Initial Pressure" or "Therapy Pressure" , press 
and rotate dial to set suitable value. 

 

y

You can rotate the dial to the right or to the left to rise or decrease therapy pressure (each step is 0.5 
cmH

2

O / hPa).

 

y

After setting therapy pressure, press dial to save your change.

 

y

Highlight the "Back", press dial to back to home page.

 

y

You can set "Max Pressure" , "Min Pressure" , "Initial Pressure" and "Therapy Pressure" of APAP 
mode according to above method.

Notice

When you try to make the max pressure lower than min pressure,initial pressure or therapy pressure, 
these pressures will be equal to the max pressure.

③ Set Pressure Unit

 

y

At clinical page, highlight "Unit".

 

y

Rotate the dial to the right or left, set your pressure unit (cmH

2

O or hPa).

 

y

After setting pressure unit, highlight the "Back" , press dial to back to home page.

④ Set FPS Level

 

y

At clinical page, Highlight "FPS Level".

 

y

Rotate the dial to the right or left, set your level of FPS (0-3 level).

 

y

After setting level of FPS, highlight the "Back", press dial to back to home page.

⑤ Set Ramp

 

y

At clinical page, Highlight "Ramp".

 

y

Rotate the dial to the right or left, set your ramp time (0-45minutes, step value is 5 minutes).

 

y

After setting Ramp, highlight the "Back", press dial to back to home page.

Summary of Contents for YH-550

Page 1: ...hnology Co Ltd Add of manufacturer No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 Web address www yuyue com cn Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany IFU V 09 version 01 Prepared date 2019 10 1030400 0A Paper ...

Page 2: ...19 14 Traveling 20 15 Trouble shooting 20 16 Specified of technology 23 17 Symbols 27 18 Limited warranty 28 19 Repairing 29 20 List of cables 29 21 Technical description 29 22 Warranty card 34 Content Welcome The YH 550 is YUWELL s auto adjusting pressure device The YH 350 is YUWELL s continuous positive airway pressure CPAP device Warning Read this entire guide before using the device Caution In...

Page 3: ...angulation y y If you notice any unexplained change in the performance of the device if it is making unusual sounds if the device or the power supply are dropped or mishandled or if the enclosure is broken discontinue use and power off the equipment then contact your provider or YUWELL Service Center y y Do not place the device where it can be bumped or where someone is likely to trip over the pow...

Page 4: ...ies with device Non YUWELL parts may reduce effectiveness of the treatment and or damage the device y y Use only vented masks recommended by YUWELL or by your doctor with this device Fitting the mask without the device blowing air can result in rebreathing of exhaled air Make sure that the mask vent holes are kept clear and unblocked to maintain the flow of the fresh air into the mask y y Do not u...

Page 5: ...ter tank 1 Air filter 2 Air tube 1 SD card 1 Power adapter and Cable as part of the main device 1 User manual 1 Bag 1 7 Picture and explanation for product Air outlet Dial START STOP button Interface silicone cover Air filter Indicator light the colour is blue and it will light with the backlight of LCD lighting Screen Movable cover component Knob cover component Air outlet of main device to water...

Page 6: ...ifficult to operate the disconnection device The following section will help you install your device by yourself 1 2 1 Place the device on a stable level surface Rotate the knob cover component from the to 2 Pull the knob cover component out 3 4 3 Press the water tank gently and pull it away from the device 4 Open the movable cover component 5 6 5 Fill the distilled water into the water tank pleas...

Page 7: ... the device is delivering gas pressure the heating plate is heating including the situations that the screen shows Large Leakage and the backlight of the screen is off When the level of the humidity is 0 the heat plate is not heating Also if the device is not delivering gas pressure the heat plate will not heat either Ramp 15 Humidity 6 APAP 6 0 15 0 6 0 Pressure Picture 1 therapy f f 10 2 Stop th...

Page 8: ...urn up the humidity If you are getting any moisture in your mask turn down the humidity You can set Humidity level between 0 to 6 where 0 means this function is disabled 1 is the lowest humidity setting and 6 is the highest humidity setting The temperature of each level on the heater plate is as following accuracy is 4 0 OFF 1 33 2 35 3 40 4 45 5 50 6 55 This temperature has been tested with empty...

Page 9: ...our change y y Highlight the Back press dial to back to home page Set Pressure y y At clinical page choose CPAP mode Highlight Initial Pressure or Therapy Pressure press and rotate dial to set suitable value y y You can rotate the dial to the right or to the left to rise or decrease therapy pressure each step is 0 5 cmH2O hPa y y After setting therapy pressure press dial to save your change y y Hi...

Page 10: ...4 0 4 0 Initial Pressure Picture 6 1 Clinical menu CPAP Picture 6 2 Clinical menu APAP 12 Caring for your device It is important that you regularly clean your device to make sure you receive optimal therapy The following sections will help you with disassembling cleaning checking reassembling your device f f 12 1 Disassembling 1 2 1 Pull the power connector out of device 2 Hold the cuff of the air...

Page 11: ... tank and air tube thoroughly and allow to dry out of direct sunlight and or heat Temperature should not exceed 40 Warning It is generally not necessary to sterilize your water tank if you are following the cleaning method When the water tank is contaminated it can be disinfected by immersing the water tank in a water which temperature is 75 2 for 30 minutes f f 12 3 Checking You should check powe...

Page 12: ... tube firmly onto the assembled mask 13 Therapy data BreathCare PAP records your therapy data for you and your care provider so they can view and make changes to your therapy if required The data is recorded and then transferred to your care provider via a SD card 1 Open the silica gel stopper at the left side of the device 2 Push the SD card into the SD card connector 3 Remove the SD card from th...

Page 13: ...ack After therapy start the screen backlight went off In other case the power do not connect firmly Press dial to turn on the screen light Check the connection of power ensure it connect with device firmly My water tank is leaking Water tank assemble incorrectly or it is broken Check the water assemble correctly Contact your provider if your water tank is broken f f 15 2 Other trouble Message in L...

Page 14: ...evel Sound pressure level measured according to ISO 80601 2 70 2015 CPAP mode BreathCare PAP 32dB A Sound power level Sound power level measured according to ISO 80601 2 70 2015 CPAP mode BreathCare PAP 40dB A Physical properties Dimensions length width height 272 5mm 181mm 93mm or 10 73 7 13 3 7 Weight About 1419 g Air tube Plastic hose about 1 8m Maximum volume of water tank 260 10 mL Materials ...

Page 15: ...ormance Humidification system output 12mg L Relative humidity 50 Humidification system Pressure drop ISO 5367 2014 Flowrate L min Pressure drop cmH2O 30 0 18 60 0 6 90 1 2 Gas leakage at the maximum operating pressure ISO 5367 2014 2 L min Displayed values Value Range Display resolution Pressure measurement Mask pressure 4 20 cmH2O 0 1 cmH2O Value Accuracy Pressure measurement Mask pressure 2 of t...

Page 16: ...ean Authorized Representative Environmental information EU directive 2012 19 EE Waste Electrical and Electronic Equipment WEEE Protected against finger sized objects and against dripping water on the surface of the device f f 17 2 Disposal statement Warning Please contact local authorities your provider or YUWELL to determine the proper method of disposal of this device 18 Limited warranty YUWELL ...

Page 17: ...issions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC61000 3 3 Complies f f 21 2 Compliance information for Immunity test Immunity test Compliance level Electrostatic discharge ESD IEC61000 4 2 8kV contact 15kV air Electrical fast transient burst IEC61000 4 4 2kV for power supply lines Surge IEC61000 4 5 1kV differential mode Voltage dips shor...

Page 18: ...m The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alternative to FM modulation 50 pulse modulation at 18 Hz may be used because while it does not represent actual modulation it would be worst case f f 21 3 Precautions According to IEC60601 1 2 2014 Brea...

Page 19: ...ice organizations that have not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident act of god or human factor Product which is not involved in quality warranty sheet User sign Date Contact Department User Add Diagnose Tel BreathCare PAP Warranty Card Model SN Invoice number Purchasing date Dealer Feedback Couplet 22 Warranty card ...

Page 20: ... damage caused as a result of improper use abuse modification or alteration of the product Repairs carried out by any service organizations that have not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident act of god or human factor Product which is not involved in quality warranty sheet User sign Date Customer Couplet ...