5.2-8.6mmol/L
CTRL 1
5.2-8.6mmol/L
CTRL 1
8.8-13.8mmol/L
CTRL 2
5.2-8.6mmol/L
CTRL 1
8.8-13.8mmol/L
CTRL 2
8.8-13.8mmol/L
CTRL 2
When “ “ and “#” are displayed simultaneously, the result is from a
control solution test, and it will not be included in the 7, 14, 30, 60, or 90
day average. When you access your test history, the control solution
results are not displayed.
If the control solution results are outside of the reference range:
After checking all of the conditions above, repeat the quality control test
with a new test strip. If the quality control test results are still outside of
the range printed on the test vial (or foil pouch), there may be a problem
with your meter. Please seek help and contact your dealer.
Two levels of control solution are available labeled Control Solution 1 and
Control Solution 2. Control Solution 1 is sufficient for most self testing
needs. If you think your meter or strips may not be working correctly, you
may also want to do a level 2 test. The ranges for both (CTRL 1 and CTRL
2) are displayed on the test strip vial (or on the foil pouch). Simply repeat
Steps 4 through 6, using Control Solution 2. For confirmation of results,
Control Solution 1 tests should fall within the CTRL1 range, and Control
Solution 2 tests should fall within the CTRL2 range. If the control solution
test results do not fall within the respective ranges, DO NOT use the
system to test blood, as the system may not be working properly. If you
cannot fix the problem, contact your dealer for help.
Please contact your dealer to purchase control solution. Each box of
control solution contains both Control Solution 1 and Control Solution 2.
Confirm you are matching the correct range. Control Solution 1
results should be matched to the CTRL1 range printed on the test
strip vial (or foil pouch).
Check the expiration date of the test strip and control solution. Make
sure that the packages have not been opened for more than 6
months. Discard any expired test strips and control solution.
Confirm that you are testing within the correct temperature range
(15-30
°
C
).
Make sure that the test strip vial and control solution bottle have
been tightly closed.
Make sure that you are using the correct brand of control solution.
Make sure you are following the user guide instructions properly.
Shake the control solution bottle, gently squeeze out the control
solution, discard the first drop, and drop the second drop onto a clean
nonabsorbent surface. Now touch the second drop to the sample
area of the test strip. Do not let the bottle come into contact with the
test strip. If the audio is on, the meter will beep, telling you that
enough control solution has been applied.
Tips:
After applying enough control solution, the display will count down 5
After the test has been completed, slide the test strip ejector to pop
out the test strip, and the glucose meter will automatically turn off.
If a big bubble forms, wipe with clean cotton paper, and then do the
following steps.
If one drop of control solution does not fill the test window, please
add another drop within 3 seconds. Otherwise discard the test strip,
and try again.
seconds and the control
solution test result will be
displayed on the screen. If the
control solution test result falls
within the control range that is
printed on the package (CTRL
1 on test strip vial or foil
pouch), this indicates that the
meter is working normally, and
the system is functioning
properly.