YSI YSI 5905 Instruction Manual Download Page 1

YSI 5905/5010
BOD Probe
Instruction
Manual

Summary of Contents for YSI 5905

Page 1: ...YSI 5905 5010 BOD Probe Instruction Manual...

Page 2: ......

Page 3: ...HANGING MEMBRANE CAPS FIGURE 2 3 WARM UP 5 ELECTRODE CLEANING FIGURE 3 5 OPERATING PROCEDURES AND PRECAUTIONS 7 PROBLEM SOLVING 9 TROUBLE SHOOTING GUIDE 9 SERVICING 11 WARRANTY AND REPAIR 13 CLEANING...

Page 4: ...ii...

Page 5: ...meters except for the Model 55 or Model 95 It must be used with an adapter 5011 in order to be used on the Model 5000 and 5100 The 5010 probe is specifically designed for use with the Model 5000 and 5...

Page 6: ...ectrolytic cell and acts as a reference electrode Upon entering the cell through the membrane oxygen is reduced at an applied potential of 0 8 V referenced to the silver electrode The reduction curren...

Page 7: ...aight out 2 Unscrew the old membrane cap from the probe Before installing a new membrane clean the probe tip with deionized water in order to remove any contaminants 3 Hold the membrane cap and fill i...

Page 8: ...4 Fill Membrane Cap with Solution Unscrew Cap Remove Stir Paddle Screw Cap on moderately tight Replace Stir Paddle Figure 2 Membrane Installation on the 5905 5010 Probe...

Page 9: ...gold cathode must be textured properly It can become tarnished or plated with silver after extended use The gold cathode can be cleaned by using the adhesive backed sanding disc provided in the 5906 M...

Page 10: ...Silver Anode It is normal for a dark layer of silver chloride to cover the silver anode After prolonged use it may become necessary to clean the anode Soak the probe in a 14 ammonium hydroxide solutio...

Page 11: ...ling through it e g nitrogen argon The meter should read less than 1 dissolved oxygen in either of these environments If it does not change the membrane or clean the probe Membrane Life Membrane life...

Page 12: ...and strong solvents that may attack the probe Probe materials that may be damaged by these substances include FEP Teflon EPR rubber ABS plastic and stainless steel Erroneous Readings Erroneous reading...

Page 13: ...ibrate Allow approximately 30 min for probe to polarize when first turned on Change membrane Clean probe see instructions under Electrode Cleaning Does not return to calibration value Allow sufficient...

Page 14: ...ange membrane Clean probe see instructions under Electrode Cleaning Motor not working Check power supply Manually turn the stir paddle to help start the motor Replace motor Loud motor Check if stir pa...

Page 15: ...he following steps 1 Remove stir paddle by pulling it straight out 2 Remove the two screws A on the outside of the housing using a 3 32 Allen wrench 3 Loosen the top nut of the cable strain relief B f...

Page 16: ...12 Housing Assembly Motor Assembly Probe Assembly Stir Paddle A B A D D C C Figure 4 Exploded View of 5905 5010 Probe...

Page 17: ...placement Limitation of Warranty This Warranty does not apply to any YSI product damage or failure caused by i failure to install operate or use the product in accordance with YSI s written instructio...

Page 18: ...CE CENTERS For the nearest authorized service center contact YSI Technical Support 1725 Brannum Lane Yellow Springs Ohio 45387 Phone 1 937 767 7241 800 897 4151 US Fax 1 937 767 9353 E Mail environmen...

Page 19: ...withhold service until appropriate cleaning decontamination and certification has been completed We will contact the sender for instructions as to the disposition of the equipment Disposition costs wi...

Page 20: ...age 5 Cleaning must be completed and certified on any product before returning it to YSI PACKING INSTRUCTIONS 1 Clean and decontaminate items to insure the safety of the handler 2 Complete and include...

Page 21: ..._______________ State ______ Zip ________ Country __________________ Model No of Device ______ Lot Number _________ Contaminant if known _____________________ Cleaning Agent s used ___________________...

Page 22: ...solved Oxygen Accuracy 0 1 mg L or 2 of reading whichever is greater Operating Temperature Range 15 to 35 C Temperature Accuracy 0 2 C Polarizing Voltage 0 8 volts Response Time Typical response to a...

Page 23: ...cludes an adapter to fit 5 taper per side smaller bottles with built in overflow funnels 5905 K 5010 K Includes a s 19 funnel 5905 L 5010 L Includes a s 24 funnel 5905 W 5010 W 6 long probe for use wi...

Page 24: ...10593 110592 110591 N A 073035 073047 005011 5905 L 059880 059902 115904 110594 110591 059911 073035 073047 005011 5905 K 059880 059902 115904 110594 110591 059933 See dealer See dealer 005011 5905 J...

Page 25: ...1725 Brannum Lane Yellow Springs Ohio 45387 USA 059882 Web 937 767 7241 800 765 4974 A59882H Fax 937 767 9353 September 99 Info YSI com www YSI com 1997 YSI Incorporated...