YSI PRO20 Instrument Operating Procedure Download Page 1

IOP 011C.3 

Effective Date: 1/09/2017 

Replaces IOP 011C.2 

Page 1 of 6 

 

 

 

 

 

 

 

 

 

U.S. Fish and Wildlife Service 

 

 

 

 

 

 

 

Marquette Biological Station 

 

 

 

 

 

 

 

3090 Wright Street 

 

 

 

 

 

 

 

Marquette, Michigan 49855 

 

 

 

 

 

 

 

U.S.A. 

 

 

 

 

 

 

 

 

 

and 

 

 

 

 

 

 

 

 

U.S. Fish and Wildlife Service 

 

 

 

 

 

 

 

Ludington Biological Station 

 

 

 

 

 

 

 

5050 Commerce Drive 

 

 

 

 

 

 

 

Ludington, Michigan 49431 

 

 

 

 

 

 

 

 

 

and 

 

 

 

 

 

 

 

 

Fisheries and Oceans Canada 

 

 

 

 

 

 

 

Sea Lamprey Control Centre 

 

 

 

 

 

 

 

1219 Queen Street East 

 

 

 

 

 

 

 

Sault Ste. Marie, Ontario P6A 2E5 

 

 

 

 

 

 

 

Canada 

 

 

INSTRUMENT OPERATING PROCEDURE

 

 

INSTRUMENT: 

 

 

Dissolved oxygen meter 

 

MODEL: 

 

 

Pro 20 

 

MANUFACTURER: 

 

 

Yellow Springs Instrument Co., Inc. 

 

PRECAUTIONS: 

 

 

POTENTIAL INTERFERENCES 

 

Chlorine, sulfur dioxide, nitric oxide, and nitrous oxide can affect readings by behaving like oxygen at 

membrane.  Static electricity may affect readings. 

Gold Cathode - The gold cathode on the sensor must be textured properly.  It may becomes tarnished or 

plated with silver after extended periods of use.  

Bubbles under the membrane cap. 

Remove excessive water in calibration/storage sleeve. 

 

SAFETY 

 

No Special precautions 

 

 