Settings
MultiLab 4010-1W
70
ba76141e07
08/2018
9.4
Cond measurement settings
9.4.1
Settings for IDS conductivity sensors
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
<ENTER>
key. After completing the set-
tings, switch to the measured value display with
<M>
.
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (
4310
,
4320
).
Default settings are printed in
bold
.
General setting
menu for
conductivity
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of
the last calibration.
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the last calibration
records (max. 10)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data stor-
age to the interface
Calibration
/
Calibration interval
1 ...
150
...
999 d
Calibration interval
for the IDS
conductivity sensor (in days).
The meter reminds you to cali-
brate regularly by the flashing
sensor symbol in the measuring
screen.
Reset
-
Resets all sensor settings to the
delivery condition (see section
9.6.1 R