YSI 5200A
135
YSI 5200A
134
-
Sensor relay menu displays configured local relay name at relay field once a
local relay has been configured at the menu being viewed.
-
Remote instrument and relay names - When Device Address is anything other
than Local Machine all relay assignment menus remain generic. This means,
Master and Slave instruments do not display their configured names, but rather
the generic names of “Master” and “Slave 1-31.”
Password
Menus → System → Password
Enable and set the password in the Password menu. Once the password function
is enabled and saved, the 5200A system menus are not accessible without first
entering the correct password at the Run Screen. The password can contain up to
12 characters.
Configuring the 5200A
1. Enter Password menu
Menus → System → Password
2. Use
▲
and
▼
to scroll and highlight submenus.
3. Press
to select.
4. Enable and configure submenus.
5. Press
to save configuration and return to Run Screen.
Notes:
-
is not password protected.
-
icon displays at Run Screen when password function is enabled.
Power Fail Menu
Menus → System → Power Fail
In the event of AC power failure, an AC version instrument’s power switches to
DC if properly wired. The switch from AC to DC power is automatic and does not
require 5200A configuration. Enable Power Fail AC Monitor to generate event when
AC power fails and is restored.
Disabling the Power Fail System will disable AC power events and disable the General
Alarm in the event of low battery for DC power.
Disable
Power Fail AC Monitor
when powering 5200A by DC power source.
Enable Low Battery Monitor to trigger the General Alarm in the event DC voltage is
≤ to low battery trigger. The Low Battery Alarm Hold Off time is five (5) seconds and
cannot be user configured. Once the low battery general alarm has been acknowl-
edged by pressing one of 4 softkeys, it is not reset. See General Alarm - page 128.
Configuring the 5200A