YSI 5200A
129
YSI 5200A
128
B
C
D
E
F
G
H
K
L
D
o not c
onnec
t or disc
onnec
t IDC r
ibbon cable
when 5200A is po
w
er
ed
. S
er
ious damage can oc
cur
.
4. Proceed with uploading firmware with the Firmware Updater utility.
See
Update 5200A Firmware - page 123.
General Alarm
Menus → System → General Alarm →
The General Alarm triggers the buzzer, general alarm relay, general alarm icon, and
5200A emails. Sensor alarms, aux alarms, slave failure (networks), and low battery
alarms trigger the General Alarm when enabled.
If individual system alarms are
not enabled, the General Alarm will not trigger.
The General Alarm (buzzer, relay, emails, icon) remains active until one of the four
softkeys
on the front panel is pressed. Pressing one of
the four softkeys while in the Run Screen stops the buzzer, de-energizes the General
Alarm relay, stops 5200A emails, and removes the General Alarm icon. Pressing
one of these keys also resets the sensor and aux alarm systems. If the alarm condi-
tion still exists after Alarm Hold period(s), the General Alarm triggers again. Low
battery and slave failure alarms are not reset when softkeys are pressed.
Sensor alarm values set the acceptable control range. When a sensor reports a value
≤ low alarm value or ≥ high alarm value, the alarm system becomes active after
the alarm hold off period. See Sensor Setup Controls and Alarms - page 149 for
additional information. If an alarm system is active and alarm relay is enabled, the
alarm relay energizes. See Aux Sensor Setup - page 157 for important aux control
and alarm system information.
Notes:
-
The Run Screen must be displayed during normal operation. All alarm
functions (5200A alarm emails, energizing of enabled alarm relays, alarm
Configuring the 5200A
icons, and buzzer) are suspended in 5200A menus. Alarm systems are reset
when menus are exited to Run Screen.
-
The General Alarm system is disabled during systemwide events such as a
Service Hold and a Firmware Download.
-
Sensor and Aux alarms must be enabled to trigger the General Alarm.
Alarm Hold Off
Menus → System → General Alarm → Alarm Hold Off→Time Units
General Alarm Hold Off is the length of time the alarm condition must exist before
the General Alarm system becomes active.
Notes:
-
To avoid alarms for spurious events set the Alarm Hold Off > than 0.
-
Aux system alarms are triggered after the Aux Hold Off Time, and not after
the General Alarm Hold Off Time.
-
Low battery alarm has a five (5) second hold off that is not user selectable.
Relay
Menus → System → General Alarm → Relay
The General Alarm relay can be wired and configured to
any
networked output
relay. The configured relay energizes when the General Alarm system is active. See
page 86 for explanation of local versus remote instruments.
Configure General Alarm relay
Use AquaManager to configure system inputs and outputs - page 146.
1. Wire relay output device to local instrument, or any networked 5X00
instrument - see page 46.
OR
1. Wire relay output device to REM - see IOEM/REM Safety and Installation
guide>Chapter 3>Step 4>Wire REM Relay Outputs.
2. Enter General Alarm Menu
Menus → System → General Alarm>General Alarm Menu opens
NEW
Configuring the 5200A