Softkey label
Function
Access to Main Menu including:
calibration submenu
system submenu
sensor setup submenu
timers submenus
Display active-non timer control & alarm
system(s)
Access to Msg’s (Messages) Menu including:
sensor logs
event logs
calibration logs
Activate Service Hold
Exit menu and return to Run Screen saving
any menu configuration changes.
Exit menu (to previous menu) without sav-
ing configuration changes.
Back to previous menu. Changes to configu-
rations are saved at return to Run Screen.
Prompt to take next action, i.e. enter a value
at calibration menus.
Save alpha and numeric configurations and
calibrations.
Disable/Enable menu option, i.e. buzzer.
Scroll left and right in alpha and numeric
keypad screens.
To acknowledge and save sensor calibration.
Prompt to confirm “Are you sure?” message
when performing certain functions, i.e. reset
to factory and clear data log menus.
Digital input control configuration from
Sensor Setup Menu.
Table 4.1
Configuring the 5200A
Run Screen
The Run Screen displays measurements and inputs from enabled sensors and condi-
tions - figure 4.2. At power up, the Run Screen is displayed immediately after the
YSI splash screen (the YSI splash screen appears for approximately 2 seconds). Icons
displayed at the Run Screen represent individual and systemwide conditions. See
Run Screen Icons
- page 70.
Run Screen and sensor display formats are user selectable. See Display Settings -
page 122 and Sensor Setup Menus - page 146.
Notes:
-
The Run Screen must be displayed during normal operation. All alarm
functions (5200A alarm emails, energizing of enabled alarm relays, alarm
icons, and buzzer) are suspended when in 5200A instrument menus.
Alarm
systems are reset when menus are exited to Run Screen.
-
Always manually poll the 5200A after uploading configuration changes and
verify the uploaded information is retained. When configuration changes
are uploaded using AquaManager at the same time configuration changes are
being made at the instrument, the AquaManager uploads may not be saved
depending on the timing of keystrokes at the 5200A.
-
The Run Screen must be displayed during normal operation.
If the 5200A
is not at the Run Screen during AquaManager uploading, pressing
at the 5200A will only save certain configuration changes.
-
At power up, 5200A systems have a 30 second hold off allowing sensor values
to stabilize before enabling any control and/or alarm devices.
Configuring the 5200A