The CAL symbol will appear at the bottom left of the display and the large portion of the
display will show
1.91
%
(or a value set previously using Advanced Setup).
2.
Press and release the [MODE] key. The large portion of the display will show
25.0C
(or a
value set previously using Advanced Setup).
25.0C
22.7
CAL
3.
Use the [UP ARROW] or [DOWN ARROW] key to change the value to the desired new
reference temperature (the allowable range is 15°C to 25°C).
15.0C
22.7
CAL
4.
Press the [ENTER] key. The word “SAVE” will flash on the display for a second to indicate
that your change has been accepted.
5.
Press the [MODE] key to return to normal operation.
4.4 Manual
Ranging
If your application is easier to perform using a manual range that you select, the YSI Model 3100
allows you to turn off the default autoranging feature. While you are making conductivity or
temperature compensated conductivity measurements, simply press and release the [ENTER]
key. The conductivity units will flash indicating that the instrument is now in a manual range.
Each additional press of the [ENTER] key will cycle the Model 3100 to a different manual range
until you return again to autoranging. Six pushes of the [ENTER] key will cycle the Model 3100
through the five available manual ranges and return the instrument to autoranging.
10.62
mS
24.8
°C
Reference temperature
New reference temperature
Flashes to indicate
Manual range