![Ypsomed YpsoPump Quick Start Manual Download Page 18](http://html2.mh-extra.com/html/ypsomed/ypsopump/ypsopump_quick-start-manual_3621846018.webp)
18
3/8:
The screen to set the basal rate
for the first hour appears, starting at
midnight (00 – 01 h). When putting into
operation for the first time (when
programming a basal rate profile for the
first time), an hourly value, which is
set and confirmed, is applied to the
next hour as a starting value.
4/8:
Select the value specified for you
by your doctor, between 0.00 U/h and
40.0 U/h, and confirm by tapping .
After each confirmation of an hourly
value, you automatically move to the
next hour. The minimum programmable
basal rate greater than 0.00 U/h is
0.05 U/h. The basal rate values
between 0.05 U/h and 1.00 U/h can
be set in increments of 0.01 U/h. The
basal rate values between 1.00 U/h
and 2.00 U/h can be set in increments
of 0.02 U/h. The basal rate values
between 2.00 U/h and 15.0 U/h can
be set in increments of 0.1 U/h. The
basal rate values between 15.0 U/h
and 40.0 U/h can be set in increments
of 0.5 U/h.
5/8:
By swiping to the right you can
step back one hour at any time and
adjust your previous entry.