63
The BGMS is not intended for serum or plasma testing.
For capillary, venous and arterial whole blood testing
only.
Inaccurate test results may be obtained at an altitude
of more than about 3,048 meters (10,000 feet).
Severe dehydration may cause inaccurately low results.
The glucose test may be invalid if abnormally
high concentrations of the following are present:
Ascorbic acid
≥
5 mg/dL
Xylose
≥
20 mg/dL
Uric Acid
≥
20 mg/dL
If the mylife
™
Unio
™
meter and mylife
™
Unio
™
test strips
are exposed to a considerable variation in temperature,
please wait 30 minutes before measurement.
Limitations of the mylife
™
Unio
™
meter
Do not use this
meter close to strong
electromagnetic
radiation sources, in
order to avoid inter-
ference with normal
operation.
Keep the meter free
of dust, water or any
liquid.