Yosemite Y504-A User Manual Download Page 10

Maintenance

Maintenance Schedule

Unlike traditional electrical chemical sensors, optical DO sensors require low maintenance.

There are no need for frequent solution filling and calibrations.

Maintenance task

Frequency

Sensor cleaning

Every 30 days

Sensor and sensor cap inspection

Every 30 days

Sensor cap replacement

Every one year

Calibration(if required of agency)

Calibration based on required schedule

Maintenance

1)

Clean the sensor surface

Wash the outer surface of sensor with tap water, if there is

still residue, using soft brush, for some stubborn dirt, household detergent can be

added in tap water to clean.

2)

Clean the cap outer surface

Remove the protective cover , flush the dirt on the light

window of the sensor with clean water, and finally put the cover on;If you need to

wipe it, please wipe it gently with a soft cloth and do not scratch it forcefully.

Otherwise, once the fluorescent film is scratched or scratched, the sensor will not

work properly

3)Clean the cap inner surface:

If water vapor or dust gets inside the fluorescent cap, the cleaning steps are as follows:

(1) Remove the fluorescent cap

(2) Rinse the inner surface of the fluorescent cap with tap water

(3) For dirt containing fat and oil, wash it in warm water with household detergent

(4) Rinse the inner surface of the fluorescent cap with deionized water

(5) Gently dry all surfaces with a clean flannelless cloth and place in a dry place to allow water

to completely evaporate

4)

Check the cable

inspect the sensor cable if there is damage.

5)

Store the cap

Regular electrode maintenance requires cap to be stored ina

protective cover with a damp sponge and checked and watered regularly, so as to
keep the fluorescent film in a moist state for a long time.If the sensor fluorescent cap

Summary of Contents for Y504-A

Page 1: ...Y504 A Online Optical Dissolved Oxygen Sensor User Manual Yosemite Technologies Ltd V 1 0...

Page 2: ...llation 7 Part List 7 Before use 7 Sensor Installation 7 Calibration 8 Calibration Solution Preparation 8 Calibration take 2 point calibration as an example 8 Appliances and raw materials required 9 M...

Page 3: ...any fault and loss caused shall not be covered by the warranty and the manufacturer shall not bear any relevant responsibility If you have any questions please contact our after sales service departme...

Page 4: ...hase difference between the blue excitation and returned red emission is measured and the result is used to calculate DO concentration Features Membraneless no electrodes design No oxygen consumption...

Page 5: ...structure chart dimension 26x151 mm xL Y504 structure chart...

Page 6: ...RS 485 MODBUS protocols Construction 3 4 inch NPT Power Requirements DC 6 12V 5 current 50mA Temperature Sensor Type NTC Sensor OD 26mm Sensor Length 177 5 mm Cable Length 10m standard 5m 15m and 30m...

Page 7: ...should be screwed tightly to avoid moisture incursion Do not use the sensor cable to pull the sensor It is required to install sensor in a secure and stable mounting bracket Make sure power supply vol...

Page 8: ...Note If the field conditions really do not allow the sensor can be directly put into the air calibration accuracy will be slightly deviated Calibration take 2 point calibration as an example Restore u...

Page 9: ...to connect the probe to the calibration software through the computer interface and you can easily complete the calibration of the sensor according to the software prompts Appliances and raw material...

Page 10: ...ed to wipe it please wipe it gently with a soft cloth and do not scratch it forcefully Otherwise once the fluorescent film is scratched or scratched the sensor will not work properly 3 Clean the cap i...

Page 11: ...oaked in water for 48 hours before continuing to work Attention Protect the inner surface of the fluorescent cap from sun exposure Do not touch the fluorescent film with your hands Avoid applying any...

Page 12: ...UTION No data displayed on controller if sensors are connected Connection issue Reconnect sensor to controller Cable failure Contact customer service Bad sensor cap or loose sensor cap Reinstall senso...

Page 13: ...blished or unpublished terrorism civil war or any government forced Damage caused due to improper use negligence accident or caused by the improper application and installation Freight for the product...