York Survey Supply 34302 Operating Instructions Download Page 2

©York Survey Supply Centre 2018

Ref:.. \operat98\instructions 18\34302.qxp 05-01-18

1. Safety Information

• Read the following safety information carefully before attempting to operate or service the meter.

• Use the meter only as specified in this manual; otherwise, the protection provided by the meter may be impaired.

Environment conditions

1. Altitude up to 2000 metres

2. Relative Humidity 90% max.

3. Operation Ambient 0 - 40

O

C

Maintenance and Clearing

1. Repairs or servicing not covered in this manual should only be performed by qualified personnel.

2. Periodically wipe the case with a dry cloth. Do not use abrasives or solvents on this instrument.

Safety Symbols

Meter is protected throughout by double insulation or reinforced insulation.

When servicing, use only specified replacement parts.

Comply with EMC

2. General Description and Features

Thank you for selecting our Sound Level Meter. To ensure that you can get the most from it, we recommend that

you read and follow the manual carefully before use.

The unit is designed according to IEC651, ANSI S1.4 for sound level meters.

The Sound Level Meter has been designed to meet the measurement requirements for industrial safety offices

and sound quality control in various environments.

• Ranges from 35dB to 130dB at frequencies between 31.5Hz and 8KHz.

• Display with 0.1dB steps on a 4-digit LCD.

• With two weighting, A and C.

• Both AC and DC signals output is available from a single standard 3.5mm coaxial socket suitable for a 

frequency analyzer, level recorder, FFT analyzer, graphic recorder, etc...

3. Specifications

Standard Applied                                         According to IEC651, ANSI S1.4

Frequency Range                                       31.5Hz - 8KHz

Measuring Level Range                              35 - 130 dB

Frequency Weighting                                  A/C

Microphone                                                 1/2 inch Electret condenser microphone

Display                                                        LCD

Digital Display                                             4 Digits

                                                                    Resolution: 0.1dB

                                                                    Display Period: 0.5sec

Time Weighting                                           FAST (125mS), SLOW (1 sec).

Level Ranges                                              Lo: 35 - 100dB and Hi: 65 - 130dB

Accuracy                                                     ±2.0dB (under reference conditions)

Dynamic Range                                          65dB

Alarm Function                                            “OVER” is shown when input is out of range.

Maximum Hold                                            Hold readings, with decay <1dB/3 minutes.

Calibration                                                   Electrical calibration with the internal oscillator (1KHz sine wave)

AC Output                                                   0.65 Vrms at FS (full scale), output impedance approx. 600

W

DC Output                                                   10mV/dB, output impedance approx. 100

W

Power Supply                                              One 9V battery 006P or IEC 6F22 or NEDA 1604

Power Life                                                   About 50 hrs (alkaline cell)

Operating Temperature                               0 - 40

O

C (32 -  104

O

F)

Operating Humidity                                     10 - 90% RH

Storage Temperature                                  -10 - 60

O

C (14 - 140

O

F)

Storage Humidity                                        10 - 75% RH

Dimensions                                                 240(L) x 68(W) x 25(H)mm

Weight                                                         210g (including battery)

Accessories                                                 9V battery, carrying case, screwdriver, instruction manual, 3.5

j

plug.

4. Nomenclature and Functions

1. Microphone

1/2 inch Electric Condenser microphone

2. Display

Serves to display the sound pressure level (dB), over or under range

“OVER”, maximum hold data “MAX HOLD” and low battery indicator

“BT”.

dB: Sound pressure level with 0.1dB resolution.

OVER: Shown when the range setting is too high or low.

3. Power and Range Switch

• Turn power ON and select measurement range.

(Hi range = 65 - 130dB, Lo range = 35 - 100dB)

• When “OVER” is indicated, slide range switch to another range for

measurement.

4. Response and Max Hold Switch

Setting the meter dynamic characteristics (fast/slow) and maximum

value hold

S (Slow response): for comparatively stable noise measurement.

F (Fast response) : for fast varying noise.

MAX HOLD: The max hold position is used to measure the maximum level of sounds. The maximum measured

level is updated continuously.

To refresh please set the switch to “F” or “S” position to cancel existing value, then, set switch to “MAX HOLD” 

position.

5. Function Switch (A/C weighting & calibration selector)

A: A-weighting

C: C-weighting

CAL 94dB: Calibration

6. Calibration Control

can be adjusted clockwise or counterclockwise to standard 94.0dB.

7. Reset Button

Serves to reset the maximum level indication.

8. Output Jack

Standard 3.5mm 3 pole coaxial output socket.

Serves to supply AC signals and log-converted DC signals to external equipment.

Outputs:

Two outputs can be accessed through 3.5mm stereo plug refer.

DC output: Logarithmic signal. 10 mV/dB.  Impedance <100

W

AC output: approx. 0.65 Vrms corresponding to each range step. Impedance = 600

W

9. Battery Cover (on bottom)

LCD Display Description

1. Sound Pressure Level measuring value, resolution 0.1dB.

2. Measuring unit.

3. When readout is out of range.

4. MAX HOLD: Maximum Hold.

5. BT: Low battery 

indicator.

5. Calibration Procedures

1. Using an acoustic calibrator

a) Make the following switch settings.

RANGE: Hi

RESPONSE: F

FUNCT: A

b) Insert the microphone carefully into the insertion hole of

the calibrator.

c) Turn on the switch of the calibrator and adjust the CAL

screw of the instrument, until the level display indicates the

desired level.

Note: Our products are calibrated before shipment. 

Recommended calibrator cycle is six months.