YONGKANG K1 User Manual Download Page 2

2, Put the rope coarser end through its already wearing thin end part and tighten

Brief Description of Front Panel

Keys function description: in standby mode, start the key instrument into the working state,
push down this button under working state, can change the display mode.

note

The machine profile picture only for your reference use, specific in kind prevail

Parameter setup

Press start button

(>

3s

),

into parameter setup .As menu 1

1. In menu 1

When the

“*”

signal is shown on the “Alm Setup”, press the button

(>

3s

and

enter into the menu 2. Press the button

<1s

can select item

then press button

(>

3s

to set

the on/off for prompt,beep,demo and screen brightness adjustment (optional "1" ," 2 "," 3 "and"
4 "). When the

“*”

signal is shown on the “Restore”,press the button

(>

3s

and all the settings

are back to the factory settings.

Attention

Using devices outdoors or under strong light, please adjust the screen brightness to a higher
proper level for observation.

It’s better for user to choose a lower brightness to conserve battery power.

2. In menu 2

When the “*”signal is shown on the “Sounds Setup”, press the button

(>

3s

and

enter into the menu 1 ,Press the button

<1s

can select item ,then press the button

(>

3s

to setup data.choose “+”or“-” to plus or minus values.

Menu 1

Menu 2

Detailed descriptions of product functions:

1. Display Type

OLED display

2. SpO2

Measurement range

70%

100%

Accuracy

±2% on the stage of 80%

100%;

±3% on the stage of 70%

79%;

Below 70% no requirement

Resolution

1%

3. PR

Measurement range

30BPM

240BPM

Accuracy

±1BPM or ±1%

the larger one

Parameters of LED probe

wavelength

radiation power

RED

660

±

2nm

1.8mW

IR

905

±

10nm

2.0mW

This parameter can be especially useful to clinicians

.

4 Power

rechargeable lithium battery

5 Automatic standby

the product shuts off by itself when no finger is in the product about 8

seconds
6.Dimension

44mm×28.3mm×26.5mm

7. Operation Environment

Operation Temperature

5

℃~

40

Storage Temperature

-10

℃~

40

Ambient Humidity

15%

80% on operation

10%

80% in storage

Air Pressure

86kPa

106kPa

8. Declaration

EMC of this product comply with IEC60601-1-2 standard.

Classification

1. Management Class for Medical Devices:

equipment

2. Anti-electric Shock Type: Internally powered equipment
3. Anti-electric Shock Degree: Type BF equipment

Maintenance and Preservation

1. Charge batteries timely when the battery power is too low.
2. Clean the surface of fingertip oximeter before it is used to diagnose patients.
3. It would be better to preserve the product in -10~40

(14-104

) and humidity is

10%-80%.

4. It is recommended that the product should be kept dry anytime. A wet ambience might

affect its lifetime and even damage the product.

Product Accessories

1. One hang lace
2. One user manual
3. One charging line
4. One adapter

optional

Guidance and manufacture’s declaration-electromagnetic radiation-for
other EQUIPMENTS and SYSTEMS

The Pulse Oximeter is designed to be used in specified electromagnetic environment .

Users of the Pulse Oximeter must use it in the following environments.

Radiation Test

Compliance

Electromagnetic environment-guidance

RF interference

CISPR 11

Group 1

RF signal of Pulse Oximeter is simply created by

its

internal function. Therefore, its RF

interference is very low and is not likely to cause

any interference to nearby electronic equipment.

RF interference

CISPR 11

Class B

The Pulse Oximeter applies to all establishments,

including domestic establishments and those

directly connected to the public low-voltage power

supply network that supplies buildings used for

domestic purposes.

Possible Problems and Resolutions

Problem

Possible reason

Solution

SpO2 or PR

can not be

shown

normally

1. Finger is not plugged correctly

2.Patient’s Oxyhemoglobin value

is too low to be measured

1. Retry by plugging the finger

2. Try more times. If you can

make sure there is no problem in

the product, please go to hospital

timely for exact diagnosis

SpO2 or PR

is

shown

unsteady

1. The finger might not be plugged

deep enough

2. Finger is trembling or the

patient is on movement status

1. Retry by plugging the finger

2. Please remain at rest

The

Oximeter

can not be

turned on

1. Inadequate power or power off

2.The

Oximeter

might

be

damaged

1. Please charge the batteries

timely

2. Please contact with local

customer service centre

Indication

lamps

are

suddenly off

1. The product automatically shuts

off when no signal is detected in 8

seconds

2. Inadequate power

1. Normal

2. Please charge the batteries

timely

Symbols and Definitions

BF type application part

Serial number

OLED display:

start/mode key

Charging jack

OLED display modes