6.3
Replacement of fuse
CAUTION
Be sure to find the cause of the blowout and take necessary measures before replacing the fuse.
(1) Turn off the power switch.
(2) The fuse is located at the lower left viewed from the front. Pull up the cap of the fuse holder
while rotating it counterclockwise, and the cap and the fuse are detached.
(3) Replace the fuse with a new one (Fuse: 250V AC / 3.15A delay type).
(4) Then fasten the cap of the fuse holder.
3
4
5
6
7
2
ALM
AC
TEMP.
1
AC250V
T3.15A
OXYGEN ANALYZER
INPUT
OUT
S
F E
U
N
L
F0602.ai
Fuse holder (Cap)
6.4
Output characteristic
Table 1 ZX8D output characteristic
%O
2
OUTPUT (mV)
%O
2
OUTPUT (mV)
0.01
168.57
7.0
24.21
0.02
153.3
8.0
21.27
0.03
144.36
9.0
18.67
0.04
138.02
10.0
16.35
0.05
133.1
11.0
14.25
0.1
117.83
12.0
12.33
0.2
102.56
13.0
10.57
0.3
93.62
14.0
8.93
0.4
87.28
15.0
7.41
0.5
82.36
16.0
5.99
0.6
78.35
17.0
4.66
0.7
74.95
18.0
3.40
0.8
72.01
19.0
2.21
0.9
69.41
20.0
1.08
1.0
67.09
20.6
0.4238
2.0
51.82
21.0
0
3.0
42.88
22.0
-1.025
4.0
36.54
23.0
-2.005
5.0
31.62
24.0
-2.943
6.0
27.61
25.0
-3.84
E (OUTPUT: mV) = 50.74 log 21/O