<4. WIRING>
4-1
IM 11M03A02-00E
4. WIRING
In this Chapter, the wiring necessary for connection to the Model ZO21DW detector is described.
4.1 General
CAUTION
NEVER supply current to the converter or any other device constituting a power circuit in
combination with the converter, until all wiring is completed.
4.1.1
Cable Specification
Cable specifications are shown in Table 4.1.
Table 4.1
Cable specifications
Terminal name of converter
Name
Need for shields Cable type Number of wires
CELL+, CELL-
HTR TC+, HTR TC-
CJ+, CJ-
Converter signal
○
CVVS
6
HEATER
Converter heater
CVV
2
L, N,
Power supply
CVV
2 or 3 *
AO-1+, AO-1-, AO-2+, AO-2-
Analog output
○
CVVS
2 or 4
DO-1, DO-2, DO-3, DO-4
Contact output
CVV
2 to 8
AC-Z, AC-S, AC-C
Automatic Calibration unit
CVV
3
DI-1, DI-2, DI-C
Contact input
CVV
3
Note *: When the case is used for protective grounding, use a 2 cores cable.
CAUTION
• Select suitable cable O.D. to match the cable gland size.
• Protective grounding should be connected in ways equivalent to JIS D type (Class 3
grounding
), (the grounding resistance is 100 Ω or less).